SCPI Command Reference
Variable Attenuator
125
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Drift> parameter.
The <Drift> parameter corresponds to a valid
drift tolerance for the power tracking via the ALC
loop, in dB. You can use the
OUTPut[1..n]:DTOlerance? MAX and
OUTPut[1..n]:DTOlerance? MIN queries to
determine a valid range for the drift tolerance.
Example(s)
INP:WAV 1310 NM
CONT:MODE POW
OUTP:POW -25.00 DBM
OUTP:ALC:STAT ON
OUTP:DTO 0.005 DB
See Also
:OUTPut[1..n]:POWer
:OUTPut[1..n]:DTOlerance?
:OUTPut[1..n]:ALC[:STATe]
:OUTPut[1..n]:DTOlerance