Remote Control
88
FLS-2600B
General Commands
*SRE
Description
This command sets bits in the Service Request Enable Register
(SRE; initial value is 255), and enables the corresponding bit in
the Status Byte Register (STB). The command can be used to
select which events can initiate a service request.
Syntax
*SRE<wsp><value>
Parameter
The <value> parameter must be an integer between 0 and 255.
*SRE?
Description
This query returns the contents of the Service Request Enable
Register (SRE).
Syntax
*SRE?
Response
An integer between 0 and 255.
*STB?
Description
This query returns the contents of the Status Byte Register
(STB).
Syntax
*STB?
Response
An integer between 0 and 255.