*RST
Description
The *RST command sets the instrument to reset settings
(factory settings stored in ROM). Pending *OPC and *OPC?
actions are canceled. The instrument becomes idle. The *RST
command clears the Input and Error queues. The following
settings remain unchanged:
Instrument interface address
Output queue
Service Request Enable register (SRE)
Standard Event Status Enable register (ESE)
Syntax
*RST
Note
For a complete list of default parameters, see
This command uses electronic parts that have a limited life
span (non-volatile memory). You must use this command with
moderation.