Using the FLS-2600 Main Menu
14
FLS-2600
Setting the Output Power
2.
Rotate the knob clockwise or counter-clockwise until the desired
wavelength is displayed. (You can select a wavelength between
1520.00 and 1570.00 nm at a resolution of 0.01 nm.) The
Press Enter
marker starts blinking on the display.
3.
Press
ENTER
to validate the new wavelength output.
Note:
If you change the wavelength while the source is active, the new
wavelength output will not be activated until you press ENTER.
It is also possible to select a wavelength from a shortlist of wavelengths
used most often, by pressing the
Up
or
Down
button and then
ENTER
. Keep
the
Up
or
Down
button pressed continuously to accelerate the scan in the
shortlist.
To modify the shortlist of available wavelengths, see
Adding a Wavelength
to the Shortlist
on page 19 and
Deleting a Wavelength from the Shortlist
on
page 21.
Setting the Output Power
The output power mode is only accessible from the main menu (i.e., not in
the
Sweep
or
Setup
menu). To set the output power mode,
1.
Press
Power
. The
Power
marker appears on the display, indicating that
you have entered the power mode.
Figure 4-2.
Power Mode
Power
dBm
nm
Tunable
Mode
Power mode marker
Target power