Safety Information
Tunable Laser Source
5
General Safety Information
➤
The unit must be positioned in a way not to block the ventilation holes
located on each side of the unit.
➤
Any adjustments, maintenance, and repair of opened units under
voltage should be avoided and carried out only by skilled personnel
aware of the hazards involved. Do not attempt internal service or
adjustment unless another person qualified in first aid is present. Do
not replace any components while power cable is connected.
➤
Operation of any electrical instrument around flammable gases or
fumes constitutes a major safety hazard.
➤
Installation of replacement parts or modification of the unit should be
carried out by authorized personnel only.
➤
Capacitors inside the unit may be charged even if the unit has been
disconnected from its electrical supply.
AC Requirements
The FLS-2600 can operate from any single-phase AC power source
between 100 and 240 V (50/60 Hz). The maximum input current is 2 A.
Power Cable
The FLS-2600 uses an international safety standard three-wire power cable.
This cable serves as a ground when connected to an appropriate AC power
receptacle. The type of power cable supplied with each unit is determined
according to the country of destination.
Only qualified electricians should connect a new plug if needed. The color
coding used in the electric cable depends on the cable. New plugs should
meet the local safety requirements and include the following features:
➤
adequate load-carrying capacity
➤
ground connection
➤
cable clamp