Alberta E-Vent User Manual Version 1.3 June 3, 2020
30
Appendix E: Tidal Volume & FiO2 Titration Tables
Look up table (Table E-1) to be used with the device in order to set tidal volumes.
Table E-1 Tidal Volume Setting Table
Tidal Volume
Setting
Volume of air
delivered
40%
174 ml
42%
189 ml
44%
204 ml
46%
220 ml
48%
236 ml
50%
253 ml
52%
271 ml
54%
289 ml
56%
308 ml
58%
327 ml
60%
347 ml
62%
367 ml
64%
388 ml
66%
410 ml
68%
432 ml
70%
455 ml
72%
478 ml
74%
502 ml
76%
526 ml
78%
551 ml
80%
577 ml
82%
603 ml
84%
630 ml
86%
657 ml
88%
685 ml
90%
714 ml
92%
743 ml
94%
772 ml
96%
803 ml
98%
833 ml
100%
865 ml
NOTE:
Tidal Volume Setting table (Table E-1) is
only valid for Ambu-Bag Spur II REF
520211000.
NOTE:
Table E 1 Tidal Volumes were developed
using the ASL 5000 with the compliance
setting of 50ml/cm H20. Alberta E-Vent
can not be used on patients with a lung
compliance less than 10 ml/cm H2O.