Alberta E-Vent User Manual Version 1.3 June 3, 2020
11
7.
Tidal Volume
setting* (LCD 1)
-
Range: 40-100%
8.
Inspiration Time
setting* (LCD 1)
-
Range: 0.5 to 3.0 seconds
9.
Trigger Pressure
setting* (LCD 1)
-
2-5 cm H2O
10.
Mode Select
button
*
selects between the
Assist and Volume Control modes.
11.
Mode LED (Volume Control)
is for patients
who are not spontaneously requesting
breaths. The patient always receives a
breath at the control parameters up to 30
bpm.
12.
Mode LED (Assist Control)
is used for
patients who can spontaneously request
breaths. Clinician can set rate up to 18 bpm.
13.
PIP Low Limit
setting* (LCD 2)
-
0-40 cmH
2
O
14.
PEEP Low Limit
setting* (LCD 2)
-
1-15 cm H2O
15.
PIP High Limit
setting* (LCD 2)
-
10-70 cmH
2
O
16.
PEEP High Limit
setting* (LCD 2)
-
5-35 cmH
2
O
17.
Respiratory Rate Alarm
setting* (LCD 2)
-
15-35 bpm
18.
Alarm Silence
supresses audible alarm for
current alarm situation for 2 minutes. A
new alarm situation will reactivate the
audible alarm.
Hold down for 5 seconds to
suppress power off / power loss alarm.
19.
Exhale
Cycle
LED
indicates
when
machine/patient is exhaling.
20.
Inhale
Cycle
LED
indicates
when
machine/patient is inhaling
21.
Alarm LED
indicates when there is an alarm
condition. Specific alarm is identified on LCD
2 and the red remote display.
22.
Alarm Buzzer
audibly notifies attendant
there is an alarm condition. Specific alarm is
identified on LCD 2 and the red remote
display.
NOTE:
*Use Scroll Wheel to select settings and
PUSH THE SCROLL WHEEL to confirm.