10
Become familiar with the meter and the different tests by reading the instructions carefully.
The Free Chlorine, Combined Chlorine, and Total Chlorine reagents are compliant for meeting
USEPA (4500-Cl G); ISO 7393/2; and German DIN 38408 G4-2 requirements.
Observe the dip time (
as required for the test
) for accurate results.
7HVWLPPHGLDWHO\DIWHU¿OOLQJWKH
CELL
with water sample when testing for oxidizers.
Be sure the
CELL
LV¿OOHGWRFDSDFLW\PO
Rinse the
CELL
with clean water immediately after completing the test.
(DPD will stain the CELL
wall if allowed to remain in the CELL)
Just before testing, rinse the sample
CELL
with the sample water several times to get a
representative sample. (
Use deionized or distilled water for rinsing if you have a limited amount of
sample
).
Store the meter and all test materials out of direct sunlight and away from chemical storage areas.
Minimize exposure of meter and test reagents to heat above 32°C (90°F).
Dry the outside of the meter when testing is complete or before storage of the meter.
When running a DPD-1 Free Chlorine test
AFTER
a Total Chlorine DPD-3, a Total Chlorine DPD-4,
or a HR Free Chlorine test, rinsing is very important to remove residual KI, which may interfere.
Each eXact
®
Strip Micro is valid for
ONLY
one test. Discard strip after single use in regular refuse
that is inaccessible to children and pets.
Each bottle of eXact
®
Strip Micro contains, at minimum, the quantity of strips notated on the bottle.
'XHWRWKHVWULSVOLWWLQJSURFHVV\RXPD\¿QGRQHRUWZRH[WUDVWULSVWKDWDUHQRWLFHDEO\VPDOOHURU
larger in width than the normal strips in the bottle. These should be discarded. Using these strips
may give incorrect results.
The eXact
®
Chlorine Plus Meter is not compatible for use with DPD-1, DPD-3, and DPD-4 powder
pillows, tablets, or liquids available from other manufacturers. Accurate results can only be
guaranteed by using genuine eXact
®
Micro strips or reagents (
reorder information below
).
Our lab testing with the Chlorine Plus meter has shown that zeroing and measuring of the sample
normally does not require any cell cover for accurate results, except in sunlight. To obtain optimal
accuracy when testing with the meter outdoors (sunlight), use the Mixing Cap/Cell Cover when
zeroing and reading the sample.
Remove batteries when meter is not used for more than a month (Warranty Requirement).
eXact
®
Chlorine Plus Tips For Best Accuracy
eXact
®
Strip Chlorine Plus Reagent Reorder Information
N/A
N/A
0.05 - 13 ppm
0.04 - 7 ppm
0.05 - 11 ppm
0.05 - 11 ppm
1.2 - 250 ppm
0.04 - 100 ppm
0.01 - 1.5 ppm
2.4 - 300 ppm
6.4 - 8.4 pH
N/A
DESIRED VALUE
1.52 ppm
PARAMETER
No.
PART NO.
# OF TESTS DETECTION RANGE
CHEMISTRY
eXact
®
6WULS0LFURP/5HDJHQW6SHFLÀFDWLRQV)RUXVHZLWKH;DFW
®
&KORULQH3OXV3DUWQR3/86
eXact
®
Micro Carrying Case w/ foam
Mini Dilution Kit II
Bromine (DPD-1)
Chlorine Dioxide (DPD-1)
Chlorine, Free (DPD-1)
Chlorine, Combined (DPD-3)
Chlorine, Total High
Hydrogen Peroxide
Ozone (DPD-4)
Peracetic Acid (PAA)
pH
Glycine (used for Chlorine Dioxide)
PRODUCT
Ready Snap™ 3
1
2
3
4
5
6
7
8
9
10
486001
487202
486636
486633
486637
486638
486672
486648
486634
486675
486639
484014
PART NO.
480903
N/A
N/A
100
100
100
100
50
50
100
100
100
50
# OF TESTS
10
N/A
N/A
DPD
DPD
DPD
DPD
KI + Buffer
DPD + Kl
DPD + KI
Phenol Red
Glycine
ACCEPTABLE VALUE
1.46 – 1.59 ppm
If your required procedure is not listed in this manual, please see the back page for our contact information.
NOTE: Because most of our products are test strips or use reagents that have little or no hazard in the quantity sold, MSDS sheets are not supplied with the test kit.
To ensure optimal performance, store your eXact
®
kit in a cool, dry place away from excess heat
(below 90°F / 32°C), moisture, and oxidizers such as Chlorine and Bromine.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.