Hydrogen Peroxide Test Procedure
1
Remove one (1)
eXact
®
Strip Micro HP, Part No. 486648
from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the
eXact
®
Strip Micro HP, Part No. 486648
into the
CELL
and immediately press
READ
. This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle
back and forth motion (approx. 2 strokes/sec).
Remove and discard the strip after “1” on the
display disappears
. The meter will begin counting up for
100 seconds
, at the end of which, the
cursor will move across the display while the meter prepares to measure the sample. Record
result displayed (this result is automatically stored in HP6). After testing is completed, rinse cell
immediately.
DIP STRIP -
(read carefully and follow procedure closely)
5
Press the
ZERO/ON
button. The cursor will move across the display, followed by
0.00 PPM
.
The sample is ready for testing.
ZERO METER
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the
MENU
button until the display shows the parameter
HP6
.
SELECT TEST: HP6
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\ZLWK
the water sample.
FILL METER WITH SAMPLE
HP
HP6
PA
PA5
6
Dip the
eXact
®
Strip Micro PA, Part No. 486675
into the
CELL
and immediately press
READ
. This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle
back and forth motion (approx. 2 strokes/sec).
Remove and discard the strip after “1” on the
display disappears
. The cursor will move across the display while the meter prepares to measure
the sample. Record result displayed (this result is automatically stored in PA5). After testing, rinse
cell immediately.
DIP STRIP AND PRESS “READ”
5
Press the
ZERO/ON
button. The cursor will move across the display, followed by
0.00 PPM
.
Sample is ready for testing.
ZERO METER
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\ZLWK
the water sample.
FILL METER WITH SAMPLE
3
Press and re-press the
MENU
button until the display shows the parameter
PA5
.
SELECT TEST: PA5
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
1
Remove one (1)
eXact
®
Strip Micro PA, Part No. 486675
from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
Peracetic Acid Test Procedure
*
NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.