Setting the Date & Time
1)
Press the Select/Control button to display the Main Menu.
2)
Use the Scroll button to move to the Settings option, Press the Select/Control button
to confirm/select Settings.
3)
Scroll to Date/Time, Press the Select/Control button to confirm/select Date/Time.
4)
Use the Scroll buttons to enter the Set Time one number at a time. Press the
Select/Control button to confirm/select each number entered before proceeding to
the next.
5)
Use the Scroll buttons to enter the Set Date one number at a time. Press the
Select/Control button to confirm/select each number entered before proceeding to
the next.
6)
When the date and time are set, your last display will be Time & Date Set, showing
the time and date you entered. The display will revert back to the default screen.
Setting Retention Values
In the event of a reset or power outage, the following table illustrates the various product
Setting Retention Values:
Setting
Default Value on Initial
Power-Up
Setting Retained After
Pushbutton Reset?
Setting Retained After AC
Power Cycle with Battery?
Setting Retained After AC Power
Cycle without Battery?
Door Ajar Alarm On/Off
ON
ON, reverts to default
ON, reverts to default
ON, reverts to default
Door Ajar Alarm Interval Setting
1 minute
N/A
N/A
N/A
Temperature Set Point
4.0 °C
Yes
Yes
Yes
Temperature Upper Alarm Threshold
8.0 °C
Yes
Yes
Yes
Termperature Lower Alarm Threshold
2.0 °C
Yes
Yes
Yes
Temperature Display Setting
°C
Yes
Yes
Yes
Temperature Resolution Setting
1 °C
Yes
Yes
Yes
Temperature Display (Control/Glycol)
Glycol
Yes
Yes
Yes
Wi-Fi Enabled On/Off
ON
Yes
Yes
Yes
Wi-Fi Connect Alarm On/Off
ON
ON, reverts to default
ON, reverts to default
ON, reverts to default
Wi-Fi Connect Delay Setting
60 Seconds
Yes
Yes
Yes
Wi-Fi SSID
None/User Selected
Yes
Yes
Yes
Wi-Fi Password
None/User Entered
Yes
Yes
Yes
Memory Alarm On/Off
ON
ON, reverts to default
ON, reverts to default
ON, reverts to default
Memory Capacity Alarm Setting
25%
Yes
Yes
Yes
Battery Alarm On/Off
ON
ON, reverts to default
ON, reverts to default
ON, reverts to default
Battery Alarm % Remaining Setting
30%
Yes
Yes
Yes
Power Out Alarm On/Off
ON
ON, reverts to default
ON, reverts to default
ON, reverts to default
Power Out Alarm Delay Setting
10 Seconds
N/A
N/A
N/A
Time and Date
4/28/15, 4:56:01 PM
Yes - After manually set
Yes - After manually set
No, reverts to default
Data Log Interval
15 minutes
Yes
Yes
Yes
Glycol Temperature Calibration Value
-0.9 °C
Yes
Yes
Yes
Control Temperature Calibration Value
-1.2 °C
Yes
Yes
Yes
First Pumpdown Flag
TRUE
Yes
Yes
Yes