background image

 2 

COLDRINK SERIES INSTALLATION AND OPERATION GUIDE

CARTRIDGE CHANGE DETERMINATION

Replace the prefilter cartridge whenever it becomes excessively dirty (observed visually or by flow restriction). Follow the 

change instructions provided with the cartridge. Replace carbon filter cartridges when capacity is reached, or when flow 

becomes inadequate, but at least annually. All carbon filter cartridges on multiple cartridge systems should be changed at 

the same time. The system pressure gauge or accessory (sold separately) Everguard

TM

 Low Pressure Alarm (LPA) provide 

quick, simple means of determining when the filter cartridge(s) should be checked. Installed on the system outlet manifold, 

the gauge or LPA can be used to monitor both dynamic (flowing) and static (line) pressure. The cartridge(s) should be changed 

when the pressure gauge needle is in the red area while equipment is in operation, and yet the needle shows adequate line 

pressure between cycles. Observe the needle during the next equipment on cycle. If the needle registers adequate pressure, 

it can be assumed that the temporary low-pressure condition was caused by a brief power failure or other incoming pressure 

disruption.  If the low pressure condition is not due to an external condition, check/change the prefilter first. If the low pressure 

condition is still not corrected, change the carbon filter cartridges.

Figure 4 – Coldrink 1

25"

14"

INLET

3/8" FNPT

Primary 

Filter

Cartridge

Prefilter

(Inside Sump)

Figure 5 – Coldrink 2

25.59"

21.5"

INLET

3/4" FNPT

OUTLET

3/4" FNPT

Figure 6 – Coldrink 3

Outlet

3/4" FNPT

INLET

3/4" FNPT

25.75"

29.59"

FLUSHING

For maximum life, all carbon filter cartridges must be flushed for five (5) minutes at full flow before use.
ALL NEW CARTRIDGES MUST BE FLUSHED AFTER EACH CARTRIDGE CHANGE BEFORE BEING PUT INTO SERVICE!
If the cartridges are new, have been properly flushed and both needle and follower register inadequate pressure, or the LPA 

is sounding, you may be experiencing inadequate water pressure or some restriction may exist in the inlet water line. In either 

case, the incoming water pressure must be improved to receive optimum filter life.
COMPLETE CARTRIDGE CHANGE INSTRUCTIONS CAN BE FOUND ON THE FILTER CARTRIDGE LABEL.
Not for residential use. For food service applications only.

11.  Open the flushing valve and flush the cartridge(s) by 

running water through at full force for five (5) minutes. 

This will set the filtering media and purge any air and 

fines, insuring maximum filter life. See illustrations.

12.  After flushing, close the flushing valve, turn on power 

to equipment, and plug the LPA (if so equipped) into the 

electrical outlet. The system is ready for operation.

Check for compliance with state and local laws and regula-

tions. Do not use with water that is microbiologically unsafe 

or of unknown quality without adequate disinfection before or 

after the system. Systems certified for cyst reduction may be 

used on disinfected waters that may contain filterable cysts.
For installations in Massachusetts, the Commonwealth of 

Massachusetts Plumbing Code 248 CMR shall be adhered to. 

Consult your licensed plumber for installation of the system. 

This system and its installation must comply with state and 

local regulations.

DIMENSIONS

Содержание Coldrink series

Страница 1: ...use a NSF approved pipe dope or plumber tape at all connections Use a backup wrench on all fittings while connecting to avoid excessive stress on the system components 4 Install a manual shut off valve leading to the system for servicing 5 Connect minimum 3 8 water line to the Coldrink 1 System inlet Use a minimum 1 2 for Coldrink 2 and 3 Systems 6 Connect tubing to flushing valve and run to drain...

Страница 2: ...ET 3 4 FNPT Figure 6 Coldrink 3 Outlet 3 4 FNPT INLET 3 4 FNPT 25 75 29 59 FLUSHING For maximum life all carbon filter cartridges must be flushed for five 5 minutes at full flow before use ALL NEW CARTRIDGES MUST BE FLUSHED AFTER EACH CARTRIDGE CHANGE BEFORE BEING PUT INTO SERVICE If the cartridges are new have been properly flushed and both needle and follower register inadequate pressure or the ...

Страница 3: ...et water ahead of system 2 Push red relief valve Unscrew bowl Remove prefilter cartridge 3 Clean bowl with filtered water only NO SPECIAL CLEANERS 4 Feel head o ring If dry lubricate with high quality silicone lubricant 5 Install new prefilter cartridge in bowl Hand tighten bowl onto head 6 Turn on water Turn on power to equipment Everpure Carbon Filter Cartridge Change 1 Shut off power to equipme...

Страница 4: ...TEL 81 0 42 775 3015 FAX INFO EVERPURE CO JP EMAIL EVERPURE SHURFLO SOUTHEAST ASIA 390 HAVELOCK ROAD 04 01 KING S CENTRE SINGAPORE 169662 65 6795 2213 TEL FAX 65 6795 2219 FAX CSEVERPURE PENTAIR COM EMAIL OPERATING SPECIFICATIONS COMBINED MODEL NAME REPLACEMENT CARTRIDGE FLOW RATE CAPACITY GPM LPM GALLONS LITRES Coldrink System 1 MC MC 1 67 6 3 9 000 34 000 Coldrink System 1 MC2 MC2 1 67 6 3 12 00...

Отзывы: