Instruction Manual
pH 1100/ 2100
49
10.4.8
Ion calibration data reset setup program P3.6
Use this program to erase all ion calibration data.
1. Enter ion setup mode as described in steps 1-4 in
Section
10.4.1 Entering ion setup mode
(page 43).
2. Use
∆
/MI
key to scroll to program
Ion Cal Reset: P3.6.
See
Figure 155
.
3. Press
ENTER
to enter program
Ion Cal Reset: P3.6
.
4. Press
>/CAL
key to toggle between
Ion Reset: Yes
and
No
. The selection will flash. See
Figure 156
.
5. Press
ENTER
to confirm selection.
•
If you selected
Yes
, the screen flashes
Ion being
Reset, Please wait...
See
Figure 157
. The meter then
returns to setup mode.
•
If
you
select
No
, the screen flashes
Reset Aborted
and
then returns to the setup mode. See
Figure 158
.
This is the last ion setup program. Press
∇
/MR
to go to the
previous program. Press
∆
/MI
to go back to the first ion
setup program. Press
EXIT
to return to mV measurement
mode.
Mode
Ion being Reset
Please wait. . .
Ion Reset: Yes No
Select & Press ENTER
Ion Cal Reset :P3.6
Press ENT. to Select
Figure 155
Figure 156
Figure 157
Figure 158
Reset Aborted