Instruction Manual
pH 1100/ 2100
33
10.2.6
pH Calibration Due Alarm Setup program P1.4
Use this program to set a reminder when the next calibration is due.
Choose from 1 to 365 days from last calibration.
1. Enter pH/Temperature setup mode as described in steps 1-4 in
10.2.1 Entering pH/temperature setup mode
(page 30).
2. Use
∆
/MI
to scroll to program
pHCalDue Alarm: P1.4
. See
Figure 91
.
3. Press
ENTER
to enter program
pHCalDue Alarm: P1.4
.
4. Press
>/CAL
to toggle between Cal Alarm Due:
On
and
Off
. The
selection will flash. See
Figure 92
.
5. Press
ENTER
to confirm selection.
•
If you selected
On,
continue to step 6
•
If you select
Off,
pressing
ENTER
returns to the setup mode.
Press
∆
/MI
to continue to the next setup program, or press
∇
/MR
to go to the previous program
.
Press
EXIT
to return to the pH
measurement mode.
6. If you selected
On
in step 5, the display reads
Key in Cal Due
Period (1-365 days) =
. Key number of days using the numeric
keypad. If you make a mistake, use the
∆
/MI
or
∇
/MR
keys to
highlight mistake, then re-enter. See
Figure 93
.
7.
Press
ENTER
to confirm number of days to next calibration,
screen flashes
Sys Updating, Please wait...
and returns to the
setup mode. Press
∆
/MI
key to continue to the next setup
program, or press
∇
/MR
to go to the previous program
.
Press
EXIT
to return to the pH measurement mode.
Keyin Cal Due Period
(1-365 days) = 1
Figure 91
Figure 92
Figure 93
Cal Alarm Due: On
Select & Press ENTER
Off
pHCalDue Alarm : P1.4
Press ENT.
to Select