Instruction Manual
pH 1100/ 2100
28
10.1.2
mV setup submenus
Press
SETUP
to enter setup mode. Use >/
CAL
to
select mV, and then press
ENTER
. Use the
∆
/MI
and
∇
/MR
keys to scroll through submenus There are 5
submenus (programs) in the mV setup mode:
10.1.3
Ion setup submenus (pH 2100 only)
Press
SETUP
to enter setup mode from the Ion
measurement mode. Use
>/CAL
to select ion, and then
press
ENTER
. Use the
∆
/MI
and
∇
/MR
keys to scroll
through submenus. There are 7 submenus (programs)
in the ion setup mode:
mV Meas. Alarm : P2.0
Press ENT. to Select
mVCalDue Alarm :P2.1
Press ENT. to Select
Stability Setup:P2.2
Press ENT. to Select
mV Cal Data : P2.3
Press ENT. to Select
mV Cal Reset : P2.4
Press ENT. to Select
mV Measurem ent Alarm
Setup :P2.0
mV Calibration Due Alarm
Setup: P2.1
Stability Setup:P2.2
View mV Calibration
Data:P2.3
mV Calibration Reset:P2.4
-Sets low and high mV alarms
-Sets calibration reminder
alarm, select 1-365 days from
last calibration date
-Sets stability indicator criteria,
off or on (high, medium or low)
-Last mV calibration date/time
mV offset value
-Clears all mV calibration data
Ion Unit Setup:P3.0
Ion Measurement Alarm
Setup:P3.1
Ion Calibration Due Alarm
Setup:P3.2
Stability Setup:P3.3
Ion Mode Setup:P3.4
View Calibration Data:P3.5
Ion Calibration Reset:P3.6
-Selects Ion measurement
unit(ppm, ppt, mg/L, g/L,
mmol/L, mol/L)
-Sets low and high Ion
alarms
-Sets calibration reminder
alarm, select 1-365 days
from last calibration date
-Sets stability indicator
criteria, off or on (high,
medium or low )
-Selects betw een high or low
Ion resolution mode (high
has ±500 mV value, low has
±1850 mV value)
-Last Ion calibration
date/time. Calibration points
with absolute mV values
Ion slopes (from low to high
concentration)
-Clears all Ion calibration
data
see page 38
see page 39
see page 40
see page 41
see page 42
see page 43
see page 44
see page 45
see page 46
see page 47
see page 48
see page 49
Ion Unit Setup :P3.0
Press ENT. to Select
IonCalDue Alarm:P3.2
Press ENT. to Select
Ion Meas. Alarm:P3.1
Press ENT. to Select
Stability Setup:P3.3
Press ENT. to Select
Ion Mode Setup:P3.4
Press ENT. to Select
Ion Cal Data :P3.5
Press ENT. to Select
Ion Cal Reset :P3.6
Press ENT. to Select