Instruction Manual MM850402 ed. 08b
65
A1 Example
The following steps show how you can make a quick measurement using the GreenLine 4000 portable analyzer.
1. Turn
the
GreenLine 4000
on. It will begin the autozero and the autodiagnostic procedure. System error and trouble
will be displayed if encountered.
GreenLine 4000
Serial Num: 1257
Version : 1.001
! Don t insert !
! probe in chimney !
Op.: Antonio
Bat: 5.2V (74.3°C)
Autozero: 180 sec
MENU
MEASURE
–
2. Select the combustible Fuel or oil used by burner. It is possible choosing between 4 preset and 6 other available
custom types.
! Now insert !
! probe in chimney !
--- FUEL --
Natural Gas
TAG SELECTION
ANALYS.
–
3. Connect the sampling probe electrical and pneumatic terminals to the
GreenLine 4000
.
4. Insert the sampling probe in chimney or stack. Be sure for mechanical positioning and flue gas intercepting.
5. Press
[
ANALYSIS
] (F1) key to start the measurements. Select the desired data page using [
V
] and [
W
] keys. Zoom
in or out to enlarge or reduce the visualized characters using the [
X
] and [
W
] keys.
6. When measurement procedure is completed, press the [
] key to print one or more reports.
MEASURE HOLD
PRESS [PRINT]
NO. COPY 1
MODEL Standard
ESCAPE
–
FEED
7. Dismount the instrument system and switch it off. Place instruments and all accessories back in the rugged and
practical case.
O2 20.9%
CO
2
0.1%
CO 37ppm
REF O2
GRAPHIC
–
O2 20.9%
Ta 28.1C
CO2 0.1%
∆
t Calc E
CO 38p
Los------
NO 0p
λ
------
Tg TC Err
Eff------
Eff-------
REF O2
–
GRAPHIC