Instruction Manual MM850402 ed. 08b
60
Serial Port
Address
: 0
Baud-rate :
9600
ESCAPE
–
Use the [
S
] and [
T
] keys to move the cursor symbol to the desired option.
Press the [
ENTER
] key to set a new ID number or baud rate.
Press
[
ESCAPE
] (F2) to return to the previous menu.
10.3.3 Analysis
par.
By Selecting this option you can change the analyzer auto-zeroing time and CH
4
& Natural gas heating probe time (we
suggest to use the pre-set values of 180 seconds for autozero time and 60 seconds for CH
4
heating time).
Analysis par.
AutoZ. Time : 180
CH4 Heat. T. : 60
ESCAPE
–
Use the [
S
] and [
T
] keys to move the cursor symbol to the desired option.
Press the [
ENTER
] key to set a new Auto-zero time.
Press
[
ESCAPE
] (F2) to return to the previous menu.
10.3.4
P decay time
By Selecting this option, you can change the timing for leak test when using the decay method. The test has a total
period of 310 seconds and the programmable time represent the interval between the measurements.
P decay Time
P1 Time 10
P2 Time 60
P3 Time 120
P4 Time 120
ESCAPE
–
Use the [
S
] and [
T
] keys to move the cursor symbol to the desired interval.
Press the [
ENTER
] key to set new acquisition intervals.
Press
[
ESCAPE
] (F2) to return to the previous menu.