Instruction Manual MM850402 ed. 08b
31
7.2 Flue
Gas
Analysis
Connect the gas probe to the analyzer as shown in chapter 7. If the optional Autozero valve is Not installed,
DO
NOT INSERT THE PROBE IN THE CHIMNEY
.
Switch the analyzer on pressing the [
ON
] key. The display will show the following page:
EcoLine 4000
Serial no.:
91257
Version: 10.0.9
GreenLine 4000
Serial no.:
91257
Version: 10.0.9
UniGas 3000+
Serial no.:
91257
Version: 10.0.9
BTU 3000
Serial no.:
91257
Version: 10.0.9
If you press a key, or after a few seconds, this page changes automatically as follows:
if the optional Autozero valve option is NOT installed, the display will show the page below:
! Don t insert !
! probe in chimney !
Op.:
Bat: 5.2V (23.5°C)
Autozero: 60 sec
MENU
MEASURE
–
if the optional Autozero valve option is installed, the display will show the page below:
!wait for autozero !
!wait for autozero !
Op.:
Bat: 5.2V (23.5°C)
Autozero: 60 sec
MENU
MEASURE
–
This page displays the following information: operator name, battery charge level, instrument internal temperature and
the seconds to end the autozero procedure.
N
OTE
: T
HE ANALYZER MUST DRAW CLEAN AIR FROM THE AMBIENT DURING THE AUTOZERO PROCEDURE
.
O
THERWISE THE
MEASUREMENTS WILL NOT BE ACCURATE
.
It is possible to enter in “MENU” page (pressing [
F1
]) or in “MEASURE” page (pressing [
F2
]). See the appropriate
paragraphs for operations.
When the autozero procedure is finished, a beeping sound advises the user. Press the [
ENTER
] key to mute the unit.
The following page will be displayed:
! Now insert !
! probe in chimney !
--- FUEL --
Natural Gas
TAG SELECTION
ANALYSIS
–
This page allows the operator to change the fuel used and to select the Tag between the available.