5.1 SET-UP
The Model 5472 Humidifier may be operated using a water tank or directly
from a faucet without a tank.
Use Distilled or Deionized water ONLY!!
A water de-ionizing column is included, suitable for up to 100 psig.
(NOTE: The life of the deionizing column will be greatly increased by
using a carbon block water filter in line with the deionizer.) When the
deionizing resin is depleted, it will change from dark to light in color.
When it is ¾ light, it must be replaced.
5.11 Using the Water Tank
1. Attach the water de-ionizing column to a faucet.
The faucet should go to the side marked “IN”.
2. Run the outlet tube to the 5-gallon water tank.
3. Turn on the water.
The water flow rate through the purifier should not exceed 8 oz. every
25-30 seconds. Water will pass at up to 10 gallons per hour.
Deionized Water Input
Water Level Protection
Float Switch
Water Valve
Float Switch
Water Valve
Ultrasonic
Transducer
Element
Air Pump
Output Tube