5.12     Using the Tap Water System 
 

1.  Attach the water de-ionizing column to a faucet. 

The faucet should go to the side marked “IN”. 
 

2.  Turn on the water. 

Find the proper flow rate before attaching to the humidifier. 
 
The water flow rate through the purifier should not exceed 8 oz. every 25-

30 seconds.  Water will pass at up to 10 gallons per hour. 

 

NOTE 

Do not increase the flow rate!  The amount of purification that can be 
performed on the water is in direct proportion to how long it takes the 
water to flow through the column.  

 

  

Slow flow rate = highly purified water.  

                                   Fast flow rate = poorly purified water.

 

 

3.  Turn on the water. 
 

Set the flow rate to the pre-determined amount.  

DO NOT OPERATE THE 

HUMIDIFIER WITH THE FAUCET 100% OPEN.

  High flow rates will 

cause the humidifier to overfill and possibly damage the unit.  The 
humidifier consumes very little water; a low flow rate will be sufficient to 
keep the unit full. 

 

 
4.  Turn on the humidifier POWER switch.   

 

(If using with an ETS Automatic Humidity Controller, see “Humidifying” in 
the Controller manual.)  The power indicator light will turn on immediately 
and the internal air pump will begin operating.   When the basin is full, the 
automatic water level switch will activate the ultrasonic transducer.  When 
that happens, the unit will begin producing a fine mist. 
 

 
 
 
 
 
 
 
 

Содержание 5462

Страница 1: ...Ultrasonic Humidification System Model 5462 Model 5472 OPERATING MANUAL 6 30 06 www electrotechsystems com...

Страница 2: ...l 5462 03 4 2 Model 5462 Operation 04 5 0 Model 5472 06 5 1 Model 5472 Set Up 07 5 11 Using the Water Tank 07 5 12 Using the Tap Water System 09 5 2 Operation 10 5 2 1 Features 10 5 22 Operating Preca...

Страница 3: ...late producing a fine mist An air moving device forces the mist out of the humidifier through a 1 plastic tube and into the enclosure A 4 110 cfm fan is provided it should be run continuously to circu...

Страница 4: ...e the chamber If the humidifier is being used with an ETS Chamber the barb is already installed 2 Place the humidifier a minimum of 12 away from the chamber 3 Cut the 1 I D tubing to the appropriate l...

Страница 5: ...is available as 115 VAC or 230 VAC 50 60 Hz The voltage must be specified at the time of purchase The circulating fan is supplied at the appropriate voltage 115 VAC or 230 VAC 4 0 MODEL 5462 The Mode...

Страница 6: ...ist 4 Rotate the MIST INTENSITY on the POWER switch This control determines how fast the water is converted from a liquid to a mist For most applications set the control at mid point II position Rotat...

Страница 7: ...mpartment 3 Never place the humidifier inside the chamber The fans and electronics will become exposed to the wet air and will quickly fail Always operate the humidifier outside the chamber and attach...

Страница 8: ...Ultrasonic Humidifiers The Model 5472 has many features not found in the Model 5462 Figure 5 0 1 5 Gallon Water Tank for long term applications Deionizer Filtering Column 10 ft OD Polytubing Quick Co...

Страница 9: ...deionizing resin is depleted it will change from dark to light in color When it is light it must be replaced 5 11 Using the Water Tank 1 Attach the water de ionizing column to a faucet The faucet shou...

Страница 10: ...the humidifier 5 Attach the provided plastic faucet to the water tank The faucet has a O D tubing quick connect fitting attached to the outlet 6 Make sure the faucet is closed then push the O D tubing...

Страница 11: ...urified water Fast flow rate poorly purified water 3 Turn on the water Set the flow rate to the pre determined amount DO NOT OPERATE THE HUMIDIFIER WITH THE FAUCET 100 OPEN High flow rates will cause...

Страница 12: ...er in the Model 5472 3 The Model 5472 can be modified to use compressed air or an inert gas as the propulsion source instead of the air pump The 5472 can also be modified for to be a close loop system...

Страница 13: ...y 6 0 MAINTENANCE CLEANING Model 5462 5472 1 Always unplug the unit before cleaning 2 Empty the unit of all water Disconnect or remove water tank or tap water source 3 Clean the surface of the transdu...

Страница 14: ...l parts of the humidifier when it is to be stored or taken out of service for any period longer than one week Do not seal the water tank in storage Leave the top off to allow the air to completely dry...

Страница 15: ...anty becomes null and void should the equipment or any part thereof be abused or modified by the customer of if used in any application other than that for which it was intended This warranty to repla...

Страница 16: ...TANTIALLY INCREASES THE CONTENT S SUSCEPTIBILITY TO SHOCK DAMAGE DO NOT PLACE INSTRUMENTS OR ACCESSORIES INSIDE OTHER INSTRUMENTS OR CHAMBERS ELECTRO TECH SYSTEMS INC WILL NOT ASSUME RESPONSIBILITY FO...

Отзывы: