ETS 5462 Скачать руководство пользователя страница 4

1.  Install the ½” NPT x 1” Hose barb on the chamber. 
 

A.  Determine the minimum height at which the barb can be installed.  
 

I. 

Place the humidifier next to the chamber. 

 
II. 

Measure up 6” from the top of the output barb on the 
humidifier.   

 

III. 

Mark the chamber at that height.  That is the minimum height 
at which the barb can be installed.  Mounting the barb lower 
may cause the humidifier to function improperly. 

 

IV. 

For a chamber and humidifier placed on the same level 
surface, the installation height should be approximately 

18”

 

B.  Drill a .825” hole at the previously determined location and install the 

barb.  The threaded portion will go inside the chamber and be held in 
place with the ½” NPT nut.  The barb will be outside the chamber.       
(If the humidifier is being used with an  ETS  Chamber,  the  barb  is      
already installed.) 

 

2.  Place the humidifier a minimum of 12” away from the chamber.  
 
3.  Cut the 1” I.D. tubing to the appropriate length, then slide the tubing over 

each 1” barb. 

 

NOTE 

Condensation will form inside the transfer tubing when the 
humidifier is in use.  The condensation must be allowed to drain 
back into the humidifier.  If there are any slack or dips in the tubing, 
water will accumulate and block the airflow.

 

 

4.  Place the fan inside the chamber. 
 

Direct the airflow toward the humidity sensor.  A strong air current will 
prevent condensation from forming on the sensor, which will produce false 
readings.  If there is no AC power in the chamber, it may be necessary to 
drill a hole to pass the power cord to the outside (this hole must be sealed 
after the cord is installed).  The fan can run continuously for all chamber 
conditions. 
 
 
 
 
 

Содержание 5462

Страница 1: ...Ultrasonic Humidification System Model 5462 Model 5472 OPERATING MANUAL 6 30 06 www electrotechsystems com...

Страница 2: ...l 5462 03 4 2 Model 5462 Operation 04 5 0 Model 5472 06 5 1 Model 5472 Set Up 07 5 11 Using the Water Tank 07 5 12 Using the Tap Water System 09 5 2 Operation 10 5 2 1 Features 10 5 22 Operating Preca...

Страница 3: ...late producing a fine mist An air moving device forces the mist out of the humidifier through a 1 plastic tube and into the enclosure A 4 110 cfm fan is provided it should be run continuously to circu...

Страница 4: ...e the chamber If the humidifier is being used with an ETS Chamber the barb is already installed 2 Place the humidifier a minimum of 12 away from the chamber 3 Cut the 1 I D tubing to the appropriate l...

Страница 5: ...is available as 115 VAC or 230 VAC 50 60 Hz The voltage must be specified at the time of purchase The circulating fan is supplied at the appropriate voltage 115 VAC or 230 VAC 4 0 MODEL 5462 The Mode...

Страница 6: ...ist 4 Rotate the MIST INTENSITY on the POWER switch This control determines how fast the water is converted from a liquid to a mist For most applications set the control at mid point II position Rotat...

Страница 7: ...mpartment 3 Never place the humidifier inside the chamber The fans and electronics will become exposed to the wet air and will quickly fail Always operate the humidifier outside the chamber and attach...

Страница 8: ...Ultrasonic Humidifiers The Model 5472 has many features not found in the Model 5462 Figure 5 0 1 5 Gallon Water Tank for long term applications Deionizer Filtering Column 10 ft OD Polytubing Quick Co...

Страница 9: ...deionizing resin is depleted it will change from dark to light in color When it is light it must be replaced 5 11 Using the Water Tank 1 Attach the water de ionizing column to a faucet The faucet shou...

Страница 10: ...the humidifier 5 Attach the provided plastic faucet to the water tank The faucet has a O D tubing quick connect fitting attached to the outlet 6 Make sure the faucet is closed then push the O D tubing...

Страница 11: ...urified water Fast flow rate poorly purified water 3 Turn on the water Set the flow rate to the pre determined amount DO NOT OPERATE THE HUMIDIFIER WITH THE FAUCET 100 OPEN High flow rates will cause...

Страница 12: ...er in the Model 5472 3 The Model 5472 can be modified to use compressed air or an inert gas as the propulsion source instead of the air pump The 5472 can also be modified for to be a close loop system...

Страница 13: ...y 6 0 MAINTENANCE CLEANING Model 5462 5472 1 Always unplug the unit before cleaning 2 Empty the unit of all water Disconnect or remove water tank or tap water source 3 Clean the surface of the transdu...

Страница 14: ...l parts of the humidifier when it is to be stored or taken out of service for any period longer than one week Do not seal the water tank in storage Leave the top off to allow the air to completely dry...

Страница 15: ...anty becomes null and void should the equipment or any part thereof be abused or modified by the customer of if used in any application other than that for which it was intended This warranty to repla...

Страница 16: ...TANTIALLY INCREASES THE CONTENT S SUSCEPTIBILITY TO SHOCK DAMAGE DO NOT PLACE INSTRUMENTS OR ACCESSORIES INSIDE OTHER INSTRUMENTS OR CHAMBERS ELECTRO TECH SYSTEMS INC WILL NOT ASSUME RESPONSIBILITY FO...

Отзывы: