5
2. Always provide a vent on the chamber being humidified.
A chamber without a vent will cause back-pressure, which could damage
the humidifier. Also, if another pressurized source is being introduced in
the chamber, the vent will allow both the humidifier and other source to
operate properly. Without a dedicated vent, the other pressurized source
will use the humidifier as a vent. This will cause the mist to back up into
the electronics compartment.
3. Never place the humidifier inside the chamber.
The fans and electronics will become exposed to the wet air and will
quickly fail. Always operate the humidifier outside the chamber and attach
it with the supplied 1” tubing.
4. Never immerse the humidifier for filling or cleaning.
5. Clean the ultrasonic transducer frequently and thoroughly.
Any dirt or particle build-up on the transducer will cause stress to the
electronics. Once the electronics overheats and stops working, the entire
humidifier must be replaced. Frequent cleaning will allow the electronics to
work normally for many years.
6. Use distilled or deionized water only.