Model 4431T Shielded Bag Test System - Operating Manual
21
D00003 Rev. K.
7.2.3 If the Cal waveform is acceptable, click on APPROVE, otherwise readjust the voltage on the 4431
Discharge Unit and click on RE-DO. Increase the voltage by 10 volts to raise the energy reading –
decrease the voltage by 10 volts to lower the energy reading. Repeat until the Cal waveform is within
range.
7.4 TEST DISCHARGES.
You will then be asked to insert a sample, close the latch, and press START TEST.
Figure 7-13. Operator Screen requesting a sample.
7.4.1 The system is ready to perform the test discharges and capture the waveform data for this
sample, and will switch to the Preview Screen.
IF
the system is in MONITORED mode, you will be asked to create each discharge. Watch for the “HV
READY” indicator to turn green, then press the red “TEST” button on the 4431 Discharge Unit front
panel.
Figure 7-14 – Waiting for Discharge.