![ETS 4431T Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/ets/4431t/4431t_operating-manual_2434443016.webp)
Model 4431T Shielded Bag Test System - Operating Manual
16
D00003 Rev. K.
7.1 Starting a test with the ETS Model 4431 Test Manager
Make sure all cables are installed before turning on the system and starting the program. To run the
program, switch on the scope, laptop, and 4431 Discharge Unit. Switch the 4431 to AUTO, Power ON,
and HV ON. Proceed as follows:
7.1.1 From your windows desktop, double-click the ETS MODEL 4431 TEST MANAGER icon. You will
see the Operator screen shown below. Instructions are always shown in the yellow highlighted field.
Figure 7-2 – Operator Screen
7.1.2 A few manual setup steps are necessary to set up the USB devices, since the windows system
prevents software from performing this automatically.
Click on the unmarked button in the upper right corner of the screen.
Figure 7-3 – Activating the tabs.