Model 4431T Shielded Bag Test System - Operating Manual
19
D00003 Rev. K.
7.1.4 To start a new test, click on the “START TEST” button.
7.1.5 The system automatically assigns a new Test Reference Number for the test, which appears at
the upper right of the screen. (Denoted by “xxxx” in this manual)
7.1.6 Enter all desired t e s t p r o f i l e information. This is for future reference – It is not edited or
checked by the system. Make sure to study and follow your laboratory’s protocol for entering this
information. Always adhere to consistent technique for entering dates, temperatures (F vs. C), etc.
Figure 7-9. Operator screen with Test Profile data
Enter the number of samples in the SAMPLES field. Any number from 1 to 20 can be entered by
typing it in, or by utilizing the up/down arrows at the left of the field.
Enter the number of discharges in the DISCHARGES field. Any number from 1 to 6 can be
entered using the same technique.
All date fields are free-form. Dates may be entered as July 26, 2016, 7/26/2016, etc. Special
care should be taken to adhere to a single date system.
All time fields are free-form. Times should be consistently entered in either AM/PM format or 24
hour format.
Note the “DegF” button adjacent to the temperature fields. Click on this to set the system which is
in use. This does not convert or change any numbers, but is stored with the test data to identify
temperature entries.
Choose the testing range according to the energy reading expected. This sets up the
oscilloscope to a reasonable starting range. The system may optimize this if needed to provide a
better display.