4

INTRODUCTION

The IR-500 Calibrator is mainly structured by 58 mm diameter target assembly 

and dry-well temperature system controlled by a microcomputer. The external 

reference thermometer is a metal cone with good heat-conducting ability, on the 

surface of thermometer there is a oxide film with emissivity 0.95.

The IR-500 Calibrator contains a class-A PT100 sensor with a temperature control 

system and wind-cold device. The calibrator has rapid heating and cooling 

functions and is switchable between °C/°F, which can be used for different 

temperature unit calibrations.

The calibrator operates  over a range of 50 °C to 500 °C with a temperature display 

resolution of 0.1 °C.

3.GENERAL SPECIFICATIONS

Range

 

 

   

50 °C to 500 °C (122 °F to 932 °F)

Resolution  

 

 

0.1 °C

Accuracy

    

   

±1 °C below 100 °C

 

 

 

   

±2 °C from 100 °C to 200 °C

 

 

 

   

±3 °C from 200 to 500 °C)

Stability

    

   

±0.1°C at Temp<100 °C (±0.2 °F at Temp<212 °F)

 

 

 

   

±0.2 °C at 1OO °C<Temp<350 °C 

 

 

 

   

(±0.4 °F at 212 °F<Temp<662 °F)

 

 

 

   

±0.4°C at 200°C<Temp<500 °C 

 

 

 

   

(±O.8 °F at 392 °F<Temp<932 °F)

Heating Time

  

 

40 minutes to max

Cooling Time

  

 

30 minutes to 100 °C (212 °F)

Emissivity

 

 

0.95 fixed

Target Size  

 

Ø58 mm

Power 

    

 

110 volt AC, 3A or 230 volt AC (±10%), 1.5A

Dimensions  

 

114 x 180 x 233 mm

Weight

    

 

2682 grams

Содержание 822-400

Страница 1: ...User Manual IR 500 BLACK BODY CALIBRATOR Product Code 822 400...

Страница 2: ......

Страница 3: ...may result if personnel fail to observe safety precautions Before working inside the equipment turn the power off and disconnect the power cord Overhead clearance is required DO NOT place this instrum...

Страница 4: ...tchable between C F which can be used for different temperature unit calibrations The calibrator operates over a range of 50 C to 500 C with a temperature display resolution of 0 1 C 3 GENERAL SPECIFI...

Страница 5: ...sively dusty or dirty environment Maintenance and cleaning recommendations can be found in the Maintenance section of this manual The instrument operates safely under the following conditions OPERATIN...

Страница 6: ...begin normal operation If the unit fails to operate please check the power connection The heater will start operating to bring the temperature of the calibrator to the set point temperature and the di...

Страница 7: ...inlet that plugs into an IEC grounded socket CAUTION Do not plug the unit into 220V if the unit is indicated as 110V This action will cause the fuses to blow and may damage the instrument Fan The fan...

Страница 8: ...0verload alarm when the target assembly temperature 4 C higher than the set temperature indicates the heating power off ALM2 0verload alarm when the target assembly temperature 4 C higher than the set...

Страница 9: ...se the SET button and press the button the green LED on the second line will indicate the current temperature unit Press DOWN to select F press UP again to select C The chosen temperature unit will fl...

Страница 10: ...ed Pressing and holding down the button the temperature will change in 1 increments The green LED display on the front panel will indicate the actual set temperature When the set temperature changes t...

Страница 11: ...11...

Страница 12: ...12 Supplied by Electronic Temperature Instruments Ltd Easting Close Worthing West Sussex BN14 8HQ 01903 202151 sales etiltd com etiltd com 822 400 24 11 20...

Отзывы: