QUICK START

 

  

UNPACKING

Unpack the calibrator carefully and inspect it for any damage that may

have occurred during shipment. If there is shipping damage notify ETI Ltd 

immediately.

Verify that the following components are present: 

• 

Calibrator

• 

Power Cord

• 

User Manual

SET UP

Place the calibrator on a flat surface with at least 8 inches of free space around 

the instrument. The stand may be swung down to raise the front of the 

instrument from a horizontal position. Plug the power cord into a grounded 

mains outlet (Note: this calibrator has two power specifications of one model 

NO:220V/AC and 11OV/AC, please make sure the power value is suitable for the 

instrument before operation.)

Turn on the power to the calibrator by toggling the power switch on. The fan 

should begin quietly blowing air through the instrument and the controller 

display should illuminate after 3 seconds. After a brief self-test the controller 

should begin normal operation. If the unit fails to operate please check the 

power connection.

The heater will start operating to bring the temperature of the calibrator to 

the set-point temperature and the display will begin to show the actual target 

temperature.

The unit should be used with a secondary certified reference thermometer 
(E.g. ETI Reference Thermometer product code 222-055)

SETTING THE TEMPERATURE

1. 

Press "UP" or "DOWN"  to change  the set-point value.

2. 

Then press "SET". The calibrator will automatically control the target assembly 

temperature to reach the set temperature in the stated time. The red LED 

display will indicate the temperature of the target assembly. When the set-point 

temperature is changed the controller  switches the well heater on or off to raise 

or lower the temperature. The displayed well temperature gradually changes 

until it reaches the set-point temperature. The well may require 5 to 10 minutes 

to reach the set-point  depending on the span. Another 5 to 10 minutes is 

required to stabilise within ±0.1 °C of the set-point. Ultimate stability may take 15 

to 20 minutes more of stabilisation time.

Содержание 822-400

Страница 1: ...User Manual IR 500 BLACK BODY CALIBRATOR Product Code 822 400...

Страница 2: ......

Страница 3: ...may result if personnel fail to observe safety precautions Before working inside the equipment turn the power off and disconnect the power cord Overhead clearance is required DO NOT place this instrum...

Страница 4: ...tchable between C F which can be used for different temperature unit calibrations The calibrator operates over a range of 50 C to 500 C with a temperature display resolution of 0 1 C 3 GENERAL SPECIFI...

Страница 5: ...sively dusty or dirty environment Maintenance and cleaning recommendations can be found in the Maintenance section of this manual The instrument operates safely under the following conditions OPERATIN...

Страница 6: ...begin normal operation If the unit fails to operate please check the power connection The heater will start operating to bring the temperature of the calibrator to the set point temperature and the di...

Страница 7: ...inlet that plugs into an IEC grounded socket CAUTION Do not plug the unit into 220V if the unit is indicated as 110V This action will cause the fuses to blow and may damage the instrument Fan The fan...

Страница 8: ...0verload alarm when the target assembly temperature 4 C higher than the set temperature indicates the heating power off ALM2 0verload alarm when the target assembly temperature 4 C higher than the set...

Страница 9: ...se the SET button and press the button the green LED on the second line will indicate the current temperature unit Press DOWN to select F press UP again to select C The chosen temperature unit will fl...

Страница 10: ...ed Pressing and holding down the button the temperature will change in 1 increments The green LED display on the front panel will indicate the actual set temperature When the set temperature changes t...

Страница 11: ...11...

Страница 12: ...12 Supplied by Electronic Temperature Instruments Ltd Easting Close Worthing West Sussex BN14 8HQ 01903 202151 sales etiltd com etiltd com 822 400 24 11 20...

Отзывы: