background image

7

PARTS AND CONTROLS

  

The user should become familiar with the calibrator and its parts.

BACK PANEL

 

Fig 1. Back Panel Description

The back panel (

Figure 1

) consists of the power cord inlet, power switch, heater 

voltage switch, and fan.

1

  Wind input

2

  Fuse of heater

3

  Fuse of temperature control system

4

  Power Input

5

  Power switch

Power Inlet - At the rear of the calibrator is the removable power cord inlet that 

plugs into an IEC grounded socket.

CAUTION: 

Do not plug the unit into 220V if the unit is indicated as 110V. This 

action will cause the fuses to blow and may damage the instrument.

Fan - The fan inside the calibrator has two speeds and runs continuously when 

the unit is being operated to provide cooling for the instrument.

The fan runs slow for heating and maintaining operation and runs fast for rapid 

cooling. Slots are provided for airflow. The area around the calibrator must be 

kept clear to allow adequate ventilation. The airflow is directed out the front and 

can be extremely hot

Содержание 822-400

Страница 1: ...User Manual IR 500 BLACK BODY CALIBRATOR Product Code 822 400...

Страница 2: ......

Страница 3: ...may result if personnel fail to observe safety precautions Before working inside the equipment turn the power off and disconnect the power cord Overhead clearance is required DO NOT place this instrum...

Страница 4: ...tchable between C F which can be used for different temperature unit calibrations The calibrator operates over a range of 50 C to 500 C with a temperature display resolution of 0 1 C 3 GENERAL SPECIFI...

Страница 5: ...sively dusty or dirty environment Maintenance and cleaning recommendations can be found in the Maintenance section of this manual The instrument operates safely under the following conditions OPERATIN...

Страница 6: ...begin normal operation If the unit fails to operate please check the power connection The heater will start operating to bring the temperature of the calibrator to the set point temperature and the di...

Страница 7: ...inlet that plugs into an IEC grounded socket CAUTION Do not plug the unit into 220V if the unit is indicated as 110V This action will cause the fuses to blow and may damage the instrument Fan The fan...

Страница 8: ...0verload alarm when the target assembly temperature 4 C higher than the set temperature indicates the heating power off ALM2 0verload alarm when the target assembly temperature 4 C higher than the set...

Страница 9: ...se the SET button and press the button the green LED on the second line will indicate the current temperature unit Press DOWN to select F press UP again to select C The chosen temperature unit will fl...

Страница 10: ...ed Pressing and holding down the button the temperature will change in 1 increments The green LED display on the front panel will indicate the actual set temperature When the set temperature changes t...

Страница 11: ...11...

Страница 12: ...12 Supplied by Electronic Temperature Instruments Ltd Easting Close Worthing West Sussex BN14 8HQ 01903 202151 sales etiltd com etiltd com 822 400 24 11 20...

Отзывы: