ThermaChoice UBT System Service Manual
4 - 16
P/N 01035 Rev. C
13. Attach the Power Entry Cable faston connectors to the Power Entry Module (blue
wire – lower left, brown wire – lower right).
14. Install the ISO board in the ISO box base.
15. Install the ISO cover onto the ISO box base including the 2 4-40 socket head cap
screws.
16. Use the #1 Phillips screwdriver to reattach the surge arrestor cable to the MDU
chassis.
17. Install the upper half of the case, using the Oval security screwdriver to install the
four fasteners.
Power Entry Cable 00868
Warning:
Disconnect the AC power from the device before servicing. Failure to observe this
precaution could result in death or injury.
Required Tools: Oval security screwdriver (TR Fastenings, Montgomery Village, MD,
Phone 301-947-1001, Part Description: MonoDrive One), 9/64” Allen wrench, 5/64"
Allen wrench, #1 Phillips screwdriver, 3/32" Allen wrench, needle nose pliers
Caution:
The interior of the instrument is ESD sensitive. Use appropriate ESD precautionary
procedures.
Removal:
1.
Using the Oval security screwdriver, remove the four fasteners securing the upper
half of the case and remove the upper half of the case.
2.
Put controller unit on its right side.
3.
Follow the procedure described in steps 8 and 9 in the ISO Board removal section for
removal of the springs and retention pin of the circulation connector.
4.
Set controller unit on its base.
5.
Use the 3/32” Allen wrench to remove the 2 4-40 socket head cap screws that hold
the ISO box cover to the ISO box base.
6.
Remove the ISO board from the ISO box base.
7.
Disconnect the connector from the AUX board at JP3.
8.
Using the 9/64” Allen wrench, remove the 2 socket head cap screws which retain the
front bezel to the side extrusion.
9.
Using the #1 Phillips screwdriver, remove the fastener securing the Surge Arrestor
Ground and Patient Cable Ground from the MDU mounting bracket.
10.
Disconnect the Power Entry Cable from the Power Supply.
Содержание Gynecare ThermaChoice II
Страница 1: ...ThermaChoice UBT System Service Manual P N 01035 Rev C THERMACHOICE II UBT System Service Manual...
Страница 26: ...ThermaChoice UBT System Service Manual 3 11 P N 01035 Rev C Surge Arrestor ASSY CABLE SURGE ARRESTOR PN 01347...
Страница 50: ...ThermaChoice UBT System Service Manual A 1 P N 01035 Rev C Appendix A Schematics...
Страница 51: ...ThermaChoice UBT System Service Manual B 1 P N 01035 Rev C Appendix B Electrical Electronic Parts List...
Страница 52: ...ThermaChoice UBT System Service Manual C 1 P N 01035 Rev C Appendix C Operating Manual...