ThermaChoice UBT System Service Manual
4 - 10
P/N 01035 Rev. C
Note:
Ensure that the isolation washers are in place.
6.
Using the #1 Phillips screwdriver, install the grounding wires from the Patient Cable
and surge suppressor to the MDU mounting bracket. Torque the fasteners to 79 – 81
in./oz.
Note:
Due to different lead lengths, observe polarity when connecting the CCS Switch Cable to
the CCS Switch.
7.
Connect one wire of the CCS Switch Cable to the center terminal of the CCS Switch.
8.
Connect one wire of the CCS Switch Cable to the 90-degree terminal of the CCS
Switch.
9.
Connect the motor tachometer to the AUX Board JP1 connection.
10.
Install the ISO box base into the lower case using the 5 fasteners and the 5/64” Allen
wrench. Torque the fasteners to 39 – 41 in./oz.
11.
Position and install the ISO Board into the box base.
12.
Install the ISO box cover.
13.
Using a 3/32" Allen wrench, secure the two 4-40 socket head cap screws to the ISO
box.
14.
Reattach the Power Entry Cable to the Power Supply.
15.
The position of the catheter coupling on the MDU motor shaft must be set using the
procedure defined in the Controller Assembly Procedure 00953 with the 01374
alignment gage that has been calibrated using procedure 01006.
16.
Install the upper half of the case, using the Oval security screwdriver to install the
four fasteners.
Heater Fuse 01620
Warning:
Disconnect the AC power from the device before servicing. Failure to observe this
precaution could result in death or injury.
Tools needed: Oval security screwdriver (TR Fastenings, Montgomery Village, MD,
Phone 301-947-1001, Part Description: MonoDrive One), Flat blade screwdriver
Caution:
The interior of the instrument is ESD sensitive. Use appropriate ESD precautionary
procedures.
Removal:
1.
Using the Oval security screwdriver, remove the four fasteners securing the upper
half of the case and remove the upper half of the case.
Содержание Gynecare ThermaChoice II
Страница 1: ...ThermaChoice UBT System Service Manual P N 01035 Rev C THERMACHOICE II UBT System Service Manual...
Страница 26: ...ThermaChoice UBT System Service Manual 3 11 P N 01035 Rev C Surge Arrestor ASSY CABLE SURGE ARRESTOR PN 01347...
Страница 50: ...ThermaChoice UBT System Service Manual A 1 P N 01035 Rev C Appendix A Schematics...
Страница 51: ...ThermaChoice UBT System Service Manual B 1 P N 01035 Rev C Appendix B Electrical Electronic Parts List...
Страница 52: ...ThermaChoice UBT System Service Manual C 1 P N 01035 Rev C Appendix C Operating Manual...