ThermaChoice UBT System Service Manual
4 - 1
P/N 01035 Rev. C
Section Four
Repair Procedures
Assembly, AUX Board 01204
Warning:
Disconnect the AC power from the device before servicing. Failure to observe this
precaution could result in death or injury.
Required Tools: Oval security screwdriver (TR Fastenings, Montgomery Village, MD,
Phone 301-947-1001, Part Description: MonoDrive One), 5/16" nut driver
Caution:
The interior of the instrument is ESD sensitive. Use appropriate ESD precautionary
procedures.
Removal:
1.
Using the Oval security screwdriver, remove the four fasteners securing the upper
half of the case and remove the upper half of the case.
2.
Remove the connections from the AUX Board JP1, JP2, JP3 and JP4.
3.
Using the 5/16" nut driver, remove the 6 keps nuts retaining the AUX Board.
4.
Remove the AUX Board.
Replacement:
Notes:
1) Prior to the replacement of the AUX Board, confirm that 6 mounting stud spacers are
present.
2) Upon installation, position the AUX Board so that the speakers are above the holes
within the lower case.
1.
Using the 5/16" nut driver, install the 6 keps nuts. Torque 89 – 91 in./oz.
2.
Reconnect JP1-JP4 to the AUX Board. Note the routing of the ISO-Micro Cable
(Page 3-2)
3.
Install the upper half of the case, using the Oval security screwdriver to install the
four fasteners.
Содержание Gynecare ThermaChoice II
Страница 1: ...ThermaChoice UBT System Service Manual P N 01035 Rev C THERMACHOICE II UBT System Service Manual...
Страница 26: ...ThermaChoice UBT System Service Manual 3 11 P N 01035 Rev C Surge Arrestor ASSY CABLE SURGE ARRESTOR PN 01347...
Страница 50: ...ThermaChoice UBT System Service Manual A 1 P N 01035 Rev C Appendix A Schematics...
Страница 51: ...ThermaChoice UBT System Service Manual B 1 P N 01035 Rev C Appendix B Electrical Electronic Parts List...
Страница 52: ...ThermaChoice UBT System Service Manual C 1 P N 01035 Rev C Appendix C Operating Manual...