background image

CO

2

 concentration, parts per million (ppm). NDIR CO

2

 sensor, display up to 9999ppm

DISPLAY , KEYPAD, SPECIFICATION

1

3

1

2

Relative humidity

.

Range 0.1~99.9%RH

 

3

4

Three color light bar. The status of measured CO

2

.

Default:

    

<800ppm 

     

800~1000ppm

   

 

 

>1000ppm

CO

2

 and PM2.5 alarm buzzer

Buzzer is default as OFF and can be activated through keypad

5

Ambient temperature

C/F switchable. Range -10.0~60.0C

(

14.0~140.0F)

SET

-Setting modes. Long press to enter setting mode

MUTE /+

-While buzzer sounds, short press to mute 
-While in setting mode, press to adjust value

Measuring range

o

o

-10.0~60.0 C (14.0~140.0 F)

CO

2

Temperature

Resolution

3

o

o

1ppm,1ug/m , 0.1 C/ F, 0.1%RH

Accuracy

CO

2

Temperature

o

o

+0.6 C/+0.9 F

0~9999 ppm,

+40ppm+3% of reading

(

400~2000ppm)

Other range is 10% of reading

0.1~99.9% RH

Humidity

o

+5%RH (at 25 C, 10~90%RH) ; 

o

+7%RH ( at 25 C , other range ) 

Humidity

Response time
CO

2

Tair

Operating

Storage 

Power supply

Meter size 

Weight

Standard package

<30 seconds (63% step change)

<2 mins

 

(90% step change)

o

0~50 C , 0~95% RH (

avoid condensation)

o

-20~ 50 C,0~95%RH(avoid condensation)

5VDC USB type C port

610g

Meter, manual, USB type C cable
,Hanging accessories.

Power consumption    

<500mA

<20 minutes ( 90% step change)

Humidity

257 x 168 x 45.5 mm (L x W x T)

SPECIFICATION

2

5

2

4

7

6

8

9

3

3

PM2.5 dust concentration, ug/m  (ppm). range  0 to 999 ug/m   

Health index, based on 24 hour PM2.5 average. Scale from 1 to 7.  1 is the best condition. 

6
7

12000 times record auto logging function

8

9

3

3

3

Three color light bar. The status of measured PM2.5

.

Default:

    

<35ug/m  

     

35~55ug/m

   

 

 

>55ug/m

-Press longer time to start or stop logging 

II

PM2.5

3

0~999 ug/m

PM2.5

PM2.5

<30 seconds(90% step change)

Auto Logging   12000 times record

3

0-100: +/-10 ug/m ;101~999:+/-10% rdg.

www.ete.co.th

Содержание 7710

Страница 1: ...rant and other public places where personal comfort and healthy are important The standard package contains below CO2 monitor x 1pc USB C cable x 1pc NOTE provided USB C cable is for both data transfer and power charging Operation manual x 1pc Sticker Screws for wall mount hanging dust filter x 1 pack MATERIAL SUPPLIED CAUTION Please take off the AC power adapter when this monitor left idle for lo...

Страница 2: ...g 0 1 99 9 RH Humidity o 5 RH at 25 C 10 90 RH o 7 RH at 25 C other range Humidity Response time CO2 Tair Operating Storage Power supply Meter size Weight Standard package 30 seconds 63 step change 2 mins 90 step change o 0 50 C 0 95 RH avoid condensation o 20 50 C 0 95 RH avoid condensation 5VDC USB type C port 610g Meter manual USB type C cable Hanging accessories Power consumption 500mA 20 minu...

Страница 3: ...healthy Hazardous 0 0 12 0 12 1 35 4 35 5 55 4 55 5 150 4 150 5 250 4 250 5 350 4 350 5 500 7 Hazardous 5 Setting Long pressing SET key to enter the settings And then press SET key to select setting modes in sequence bL Brightness of light bar ALC 2 0 CO2 Alarm color ALC 3 0 PM2 5 Alarm color Unit temperature unit Buzzer Pressure compensation and ABC auto background calibration Press key to adjust...

Страница 4: ... while still connected with computer Windows O S Download memory as csv file to computer csv file could be easily used by many analysis software Within 30 mins Within 30 mins hanging the device in the location where you want to take record Several Days II Press stop key longer time to stop the recording Double click Logger configuration tool to setup the logger The logger configuration screen offe...

Страница 5: ...uracy Press SET keys at the same time till CAL and 400 appears Long pressing SET key to start the calibration During the calibration 400 will flash on display After 5 minutes the display return to normal measuring mode to indicate the calibration is done During the calibration procedure long pressing key can quit the calibration function CAUTION Do not calibrate the meter in the air with unknown C...

Страница 6: ...oP and real time CO2 value will appear in turns Please return to computer to download data and re program again NO response at all The meter might crash after long term usage There are two methods to power on again One unplug the power cable for 2 hours and then plug again If you could not wait for 2 hours time to discharge please use any suitable and available pin to do the hardware reset The res...

Отзывы: