Long press “

Start

” key to activate the logger till “

REC”

 LED flashes. If the logger is programmed as start 

delay 5 mins, the “

REC”

 LED  will consistent display and then turn into flashing after 5 minutes.  

To stop recording before memory is full, press “

STOP

” key longer time till “

REC

” LED went off. There are 

12000 records memory. While memory is full, 

FULL

 icon will appear on display to remind user. 

Connect to computer to read out.

 

The procedure is same like connecting to setup. 

Click “

Data Download

” button on setup screen and choose where to save the logged file to. 

The read out csv report can be opened by windows excel. 

After setting, click “

Save”

 to synchronize the setting and your PC time zone to this datalogger. 

Finally, close the setup screen and disconnect logger with computer. Within 30 minutes, Install the 
logger to the place where you want to monitor and record the indoor air quality. 

The saved report contains:
Configuration Time, Start Time, 
Start delay, Sampling rate, 
ID no. and Total records number. 

Each record contains Date and 
time, PM2.5, CO

2

, TA, RH(%) 

and health index.

5

CO

2

 CALIBRATION

The meter is calibrated at standard 400ppm CO  concentration in factory. While you turn on the meter

2

for the first time, simply leave it in fresh air condition for 5 minutes to check if it is within specification. If 
it is, no need to do re-calibration. 
Nevertheless, you may still do either ABC or manual calibration regularly to maintain good accuracy.
Press 

“ SET”

 +

 “+”

keys at the same time till “

CAL” 

and

 

400

” appears. Long pressing 

“SET”

 key 

to start the calibration. During the calibration, 

“400” 

will flash on display. After 5 minutes, the display 

return to normal measuring mode to indicate the calibration is done. 
During the calibration procedure, long pressing  

“+”

 key can quit the calibration function. 

CAUTION: 

Do not calibrate the meter in the air with unknown CO  level.  Otherwise, it will be taken as 

2

400 ppm and leads to inaccurate measurement. The manual calibration is suggested to be done outdoor 
with good ventilation, fresh air and sunny day where CO  level is around 400ppm. Do not calibrate in 

2

rainy day because high humidity will affect the CO  level in air. Do not calibrate in places crowded with 

2

people or close to where exist high CO   such as ventilating outlets or fireplaces.

2

MAINTENANCE

PM2.5 SENSOR MAINTENANCE  

1. If dust adhering inside the sensor, the sensor itself has a function to remove the dust on the fan and 
    the path by rotating the built-in fan at high speed when used for a long time, but completely removes 
    all dust are not possible. Please don’t consider to do cleaning of the sensor by any method.
2. Please keep the air OUT opening unblocked. The filter installed
    in front of the air IN opening is to filter the big dust. Please 
    replace filter while necessary. 
3. Please avoid condensation inside the sensor. When inside of 
    the sensor is moisturized, this product  does not keep its 
    proper function. 
4. If the sensor is located close to noise source ( such as electric 
    dust collector, etc), the sensor output may be affected by 
    emitted noise. 

 

Air OUT

Air IN

www.ete.co.th

Содержание 7710

Страница 1: ...rant and other public places where personal comfort and healthy are important The standard package contains below CO2 monitor x 1pc USB C cable x 1pc NOTE provided USB C cable is for both data transfer and power charging Operation manual x 1pc Sticker Screws for wall mount hanging dust filter x 1 pack MATERIAL SUPPLIED CAUTION Please take off the AC power adapter when this monitor left idle for lo...

Страница 2: ...g 0 1 99 9 RH Humidity o 5 RH at 25 C 10 90 RH o 7 RH at 25 C other range Humidity Response time CO2 Tair Operating Storage Power supply Meter size Weight Standard package 30 seconds 63 step change 2 mins 90 step change o 0 50 C 0 95 RH avoid condensation o 20 50 C 0 95 RH avoid condensation 5VDC USB type C port 610g Meter manual USB type C cable Hanging accessories Power consumption 500mA 20 minu...

Страница 3: ...healthy Hazardous 0 0 12 0 12 1 35 4 35 5 55 4 55 5 150 4 150 5 250 4 250 5 350 4 350 5 500 7 Hazardous 5 Setting Long pressing SET key to enter the settings And then press SET key to select setting modes in sequence bL Brightness of light bar ALC 2 0 CO2 Alarm color ALC 3 0 PM2 5 Alarm color Unit temperature unit Buzzer Pressure compensation and ABC auto background calibration Press key to adjust...

Страница 4: ... while still connected with computer Windows O S Download memory as csv file to computer csv file could be easily used by many analysis software Within 30 mins Within 30 mins hanging the device in the location where you want to take record Several Days II Press stop key longer time to stop the recording Double click Logger configuration tool to setup the logger The logger configuration screen offe...

Страница 5: ...uracy Press SET keys at the same time till CAL and 400 appears Long pressing SET key to start the calibration During the calibration 400 will flash on display After 5 minutes the display return to normal measuring mode to indicate the calibration is done During the calibration procedure long pressing key can quit the calibration function CAUTION Do not calibrate the meter in the air with unknown C...

Страница 6: ...oP and real time CO2 value will appear in turns Please return to computer to download data and re program again NO response at all The meter might crash after long term usage There are two methods to power on again One unplug the power cable for 2 hours and then plug again If you could not wait for 2 hours time to discharge please use any suitable and available pin to do the hardware reset The res...

Отзывы: