OPERATION

1. Power on and Measure:

-Connect the USB-C cable to a PC or 5VDC wall adapter or power bank to power on the device. 
-After power on, the LED count down for 10 seconds during warm-up. 
-The device will start to measure PM2.5, CO

2

 concentration, ambient temperature and relative humidity 

 after warm-up.
-Suggest to power on for at least 30 minutes before using the logging function.

2. Light bar 

:

Besides the value display, this meter is also equipped with programmable light bar color feature.
There are 3 different light bar color for PM2.5 and CO

2

: Green, Yellow and Red. The color auto switches 

according to the measured value. Yellow color will be skipped if red light setting is identical to yellow.

3. Buzzer & How to turn off buzzer:

Buzzer function is default OFF in case of using in quiet places. It is required to activate buzzer function 
in setting mode while audible alarm is needed. While buzzer is activated and measured CO

2

 or PM2.5 is 

over alarm threshold, buzzer sounds once a second till pressing “

MUTE

” key to stop.

4. Health Index:

Long term PM2.5 value brings higher impact to human health. The health index here provides you a quick 
indication to know the average PM2.5 level of past 24 hours. This index table is defined by USA EPA. 

    

Health index  level

AQI category

3

Breakpoints (ug/m , 24H average)

1

2
3
4
5
6

Good 

Moderate

Unhealthy for sensitive groups

Unhealthy

Very Unhealthy

Hazardous

0.0~12.0

12.1~35.4

35.5~55.4

55.5~150.4

150.5~250.4

250.5~350.4

350.5~500

7

Hazardous

5. Setting:

Long pressing 

"SET"

 key to enter the settings. And then press 

"SET"

 key to select setting modes in 

sequence:  bL (Brightness of light bar), ALC 2.0 (CO

2

 Alarm color), ALC 3.0 (PM2.5 Alarm color), Unit 

(temperature unit), Buzzer,  Pressure compensation and ABC (auto background calibration).
Press 

" + "

 key to adjust the parameters.

*Note: During the setting process, the device will return to the main screen after 25 sec of idle 
time.

 

5.1 Brightness of light bar (bLt): The brightness of light bar is adjustable. From off to strong is 0 to 10, 
0 is turned off and 10 is the strongest. Default as 6. Press

 “+”

 to adjust the brightness of light bar to fit 

your need. Though light bar might be set to off, it will still blink at weakest level 1 brightness while 
measured CO

2

 or PM2.5 is higher than alarm threshold.

5.2 Alarm(ALC, 2.0 &2.1): For CO

2

, green turns into yellow, adjustable from 700ppm to 3,000ppm, 

default 800. Yellow turns into red, adjustable from 700 to 3000ppm, default 1000. Red should be larger 
than yellow, yellow color will be skipped if red light setting is intentionally identical to yellow. Interval is 
100ppm. This red light threshold is related to buzzer alarm trigger and red light bar trigger. 
To adjust alarm threshold value, Press 

“+” 

key

.

3

From 

 turns into 

Green

Yellow

Yellow 

 

turns into

flashing Red

PM2.5

CO

2

Adjustable from 700 to 3000

(default as 800)

Adjustable from 700 to 3000

(default as 1000)

Adjustable from 10 to 100

(default as 35)

Adjustable from 10 to 100

(default as 54)

www.ete.co.th

Содержание 7710

Страница 1: ...rant and other public places where personal comfort and healthy are important The standard package contains below CO2 monitor x 1pc USB C cable x 1pc NOTE provided USB C cable is for both data transfer and power charging Operation manual x 1pc Sticker Screws for wall mount hanging dust filter x 1 pack MATERIAL SUPPLIED CAUTION Please take off the AC power adapter when this monitor left idle for lo...

Страница 2: ...g 0 1 99 9 RH Humidity o 5 RH at 25 C 10 90 RH o 7 RH at 25 C other range Humidity Response time CO2 Tair Operating Storage Power supply Meter size Weight Standard package 30 seconds 63 step change 2 mins 90 step change o 0 50 C 0 95 RH avoid condensation o 20 50 C 0 95 RH avoid condensation 5VDC USB type C port 610g Meter manual USB type C cable Hanging accessories Power consumption 500mA 20 minu...

Страница 3: ...healthy Hazardous 0 0 12 0 12 1 35 4 35 5 55 4 55 5 150 4 150 5 250 4 250 5 350 4 350 5 500 7 Hazardous 5 Setting Long pressing SET key to enter the settings And then press SET key to select setting modes in sequence bL Brightness of light bar ALC 2 0 CO2 Alarm color ALC 3 0 PM2 5 Alarm color Unit temperature unit Buzzer Pressure compensation and ABC auto background calibration Press key to adjust...

Страница 4: ... while still connected with computer Windows O S Download memory as csv file to computer csv file could be easily used by many analysis software Within 30 mins Within 30 mins hanging the device in the location where you want to take record Several Days II Press stop key longer time to stop the recording Double click Logger configuration tool to setup the logger The logger configuration screen offe...

Страница 5: ...uracy Press SET keys at the same time till CAL and 400 appears Long pressing SET key to start the calibration During the calibration 400 will flash on display After 5 minutes the display return to normal measuring mode to indicate the calibration is done During the calibration procedure long pressing key can quit the calibration function CAUTION Do not calibrate the meter in the air with unknown C...

Страница 6: ...oP and real time CO2 value will appear in turns Please return to computer to download data and re program again NO response at all The meter might crash after long term usage There are two methods to power on again One unplug the power cable for 2 hours and then plug again If you could not wait for 2 hours time to discharge please use any suitable and available pin to do the hardware reset The res...

Отзывы: