MANUAL SGM700
60
Use key 1 and 2 to change the value. Key 1 is for changing the number (1-9), key
2 is for changing the position of the cursor. Confirm by pressing key 3 for >2 sec.
Increase
Decrease
Left
Right
Confirm
The following screen is visible:
In
Ind 2
you set the
No motion band
Indicator gives stable signal when weigher value is
stable within this range and time set with Ind 3.
Choose a value between: 0
–
999999. To
change the value press key 2 < 2 sec.
The following screen is visible:
Use key 1 and 2 to change the value. Key 1 is for changing the number (1-9), key
2 is for changing the position of the cursor. Confirm by pressing key 3 for >2 sec.
Increase Decrease Left
Right
Confirm
In
Ind 3
you set the
Stable time
Indicator gives stable signal when weigher value is
stable within the range set in Ind 2 and time set with Ind 3.
Choose a value between: 000.000
–
16.959 seconds. To
change the value press key 2 < 2 sec.