The test is successful: press
to go back to the work screen.
•
The test has failed: run a calibration cycle.
Calibrating the tracing table
This procedure must be followed if the above calibration test has failed.
Select
to start the cycle.
The image of the pattern holder is displayed:
Attach the pattern gauge to the pattern holder then insert it into the tracing table.
The pattern gauge is considered a left lens. For more information on fixing the gauge on the
pattern holder, refer to the section Tracing > Tracing a pattern, a demo or re-cut lens >
Mechanical tracing
(
F
p.35)
.
Select
to start the cycle.
Apply pressure on the jaws throughout the calibration cycle to ensure that the pattern holder
remains in place.
Wait: a timer
is displayed for a few minutes.
: The first calibration phase is finished.
The image of the frame gauge is displayed:
Insert the frame gauge in the tracing table.
Close the jaws.
>
>
>
1
2
3
4
5
6
U
SER
M
ANUAL
> VIII. M
AINTENANCE
&
SERVICING
119
Pro-B 300 > v1.0 - 03.19
Содержание Pro-B 300
Страница 1: ...User Manual...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 9: ...I FIRST STEPS...
Страница 17: ...II TRACING...
Страница 40: ......
Страница 41: ...III CENTERING A LENS...
Страница 69: ...IV MODIFYING THE LENS SHAPE...
Страница 78: ......
Страница 79: ...V PREPARING A DRILLED JOB...
Страница 91: ...USER MANUAL V PREPARING A DRILLED JOB 91 Pro B 300 v1 0 03 19...
Страница 92: ......
Страница 93: ...VI PREPARING LENS EDGING...
Страница 98: ......
Страница 99: ...VII TO SET UP THE DEVICE...
Страница 114: ......
Страница 115: ...VIII MAINTENANCE SERVICING...
Страница 126: ......
Страница 127: ...TECHNICAL DATA...
Страница 130: ......
Страница 131: ...GENERAL INFORMATION 9...
Страница 135: ...GLOSSARY...
Страница 139: ...TRIPOD FOR HIGH BASE LENSES Specific tripod to center high base lenses USER MANUAL GLOSSARY 139 Pro B 300 v1 0 03 19...