P
HENOMENON
& B
ASIC
EMC
STANDARD
I
MMUNITY
TEST
LEVELS
H
OME
HEALTHCARE
ENVIRONMENT
C
OMPLIANCE
LEVEL
Electrical fast transients/ bursts (IEC
61000-4-4:2012)
Input AC power port
± 2kV
100 kHz repetition frequency
± 2kV
Signal input/ output unit port
±1kV
100 kHz repetition frequency
± 1kV
Surges
Line-to-line (IEC 61000-4-5:2017)
± 1kV
Surges
Line-to-ground (IEC 61000-4-5:2017)
± 2kV
Conducted disturbances induced by RF
fields (IEC 61000-4-6:2013)
3 Vrms
0.15 MHz - 80 MHz
6 Vrms in ISM bands between
0.15 MHz and 80 MHz
80% AM at 1 kHz
3 Vrms
Voltage dips (IEC 61000-4-11:2020)
0% U
T
; 0.5 cycle
0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
0% U
T
; 0.5 cycle
0% U
T
; 1 cycle
and
70% U
T
; 25 cycle
Single phase: 0°
0% U
T
; 1 cycle
70% U
T
; 25 cycle
Voltage interruptions (IEC 61000-4-
11:2020)
0% U
T
; 250 cycle
0%U
T
; 250 cycle
U
T
is the AC supply voltage before applying inspection level.
U
SER
MANUAL
> S
AFETY
CONSIDERATION
67
AKR800 - Auto Kerato Refractometer > V4 - 05-2022 - CE0459
Содержание AKR 800
Страница 1: ...USER MANUAL...
Страница 4: ...I INTRODUCTION...
Страница 6: ...II SUPPLY PACKAGE...
Страница 8: ...III GENERAL DESCRIPTION...
Страница 19: ...IV INSTALLATION CONNECTION...
Страница 24: ...V USE OF THE DEVICE...
Страница 25: ...1 Flow operation USER MANUAL USE OF THE DEVICE 25 AKR800 Auto Kerato Refractometer V4 05 2022 CE0459...
Страница 45: ...VI SETTING OF FUNCTION ON SETUP SCREEN...
Страница 56: ...VII ERROR DISPLAY...
Страница 60: ...VIII SAFETY CONSIDERATION...
Страница 68: ...IX TROUBLESHOOTING...
Страница 70: ...X MAINTENANCE...
Страница 75: ...XI SPECIFICATIONS...
Страница 78: ...XII QR CODE...