1
2
2
1
[All]: Print out a maximum of ten data of the refractive measurement or Kerato measurement for each eye.
[Eco]: Print out only the optimum values for all of the measurement.
[Off]: Print out no data
If the red line appears at the end of the printer paper, replace the paper soon.
When [Error Printer cover opened.] is displayed, close the printer cover tightly.
The measurement values are output to the data store site set in [Terminal] if [XML and [Standard] and [Report] on
the [Export] tab in the setup is set as other than [Off].
7. Measurement of the other eye
If [R/L Auto] is set to On, the optical head automatically moves to the position measuring the opposite eye.
If [R/L Auto] is set to Off, move the optical head to the position measuring the left eye by taping the Left switch.)
Take measurements.
Perform measurements, printing out of the measurement and analysis result and external output after finishing the
measurements.
If [R/L Auto] on [Measure 2] in the setup screen are set as On, the optical head moves to the other side
automatically, and a measurement is started.
The eye to be measured cannot be switched correctly if a patient closes his/her eye or blinks during switching.
If [R/L Auto] is set to Off, press [R] or [L] switch on the opposite side.
Do not move the optical head to the other eye by tapping or holding down the screen. The device might come into contact
with a patient's nose.
8. Operation after measurement
Talk to a patient that the measurements are finished.
Tap the [Clear] switch.
All of the measurement values are deleted.
>
U
SER
MANUAL
> U
SE
OF
THE
DEVICE
31
AKR800 - Auto Kerato Refractometer > V4 - 05-2022 - CE0459
Содержание AKR 800
Страница 1: ...USER MANUAL...
Страница 4: ...I INTRODUCTION...
Страница 6: ...II SUPPLY PACKAGE...
Страница 8: ...III GENERAL DESCRIPTION...
Страница 19: ...IV INSTALLATION CONNECTION...
Страница 24: ...V USE OF THE DEVICE...
Страница 25: ...1 Flow operation USER MANUAL USE OF THE DEVICE 25 AKR800 Auto Kerato Refractometer V4 05 2022 CE0459...
Страница 45: ...VI SETTING OF FUNCTION ON SETUP SCREEN...
Страница 56: ...VII ERROR DISPLAY...
Страница 60: ...VIII SAFETY CONSIDERATION...
Страница 68: ...IX TROUBLESHOOTING...
Страница 70: ...X MAINTENANCE...
Страница 75: ...XI SPECIFICATIONS...
Страница 78: ...XII QR CODE...