ESO
SINFONI Pipeline User Manual
Doc:
VLT-MAN-ESO-19500-3600
Issue:
Issue 1.0
Date:
Date 2005-10-19
Page:
27 of 88
2. Then the user may generate a map of non linear pixels. A set of linearity raw flat field frames may be put
in the ASCII file linearity_sof.
/path_raw/LINEARITY/K/SINFO.2005-02-26T20:09:50.882.fits LINEARITY_LAMP
/path_raw/LINEARITY/K/SINFO.2005-02-26T20:10:07.455.fits LINEARITY_LAMP
/path_raw/LINEARITY/K/SINFO.2005-02-26T20:10:23.047.fits LINEARITY_LAMP
/path_raw/LINEARITY/K/SINFO.2005-02-26T20:10:38.240.fits LINEARITY_LAMP
/path_raw/LINEARITY/K/SINFO.2005-02-26T20:10:56.403.fits LINEARITY_LAMP
/path_raw/LINEARITY/K/SINFO.2005-02-26T20:11:31.128.fits LINEARITY_LAMP
/path_raw/LINEARITY/K/SINFO.2005-02-26T20:11:59.183.fits LINEARITY_LAMP
/path_raw/LINEARITY/K/SINFO.2005-02-26T20:12:27.247.fits LINEARITY_LAMP
The user can generate a non linearity bad pixel map (PRO.CATG=BP_MAP_NL) with the command:
esorex si_rec_detlin linearity_sof
This command will generate several files, including the non linearity bad pixel map, stored in the file
“out_bp_lin.fits”.
mv out_bp_lin.fits /path_pro/BP_MAP_NL_K.fits
rm -rf out*fits *.paf *.log
3. Then the user may determine the optical distortions (PRO.CATG=DISTORTION) and the slitlets dis-
tances (PRO.CATG=SLITLETS_DISTANCE).
The user will select all the files containing the string DISTORTION in their DPR.TYPE. Using those files
and the line reference table of the corresponding band (K) and a drs setup table will generate an ASCII
file distortion_sof:
/path_raw/SINFO.2005-03-14T11:28:19.007.fits FIBRE_NS
/path_raw/SINFO.2005-03-14T11:28:43.781.fits FIBRE_NS
/path_raw/SINFO.2005-03-14T11:28:59.723.fits FIBRE_NS
/path_raw/SINFO.2005-03-14T11:29:18.496.fits FIBRE_NS
.
.
.
/path_raw/SINFO.2005-03-14T11:53:22.863.fits FLAT_NS
/path_raw/SINFO.2005-03-14T11:53:45.397.fits FLAT_NS
/path_raw/SINFO.2005-03-14T11:54:08.861.fits WAVE_NS
/path_raw/SINFO.2005-03-14T11:55:10.220.fits WAVE_NS
/path_cdb/neonK.tfits REF_LINE_ARC
/path_cdb/drs_setup_wave.tfits
DRS_SETUP_WAVE
Where with
/path_cdb
we have indicated the full path to the directory containing calibration data. The
command
esorex si_rec_distortion distortion_sof