![ESO SINFONI Pipeline Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/eso/sinfoni-pipeline/sinfoni-pipeline_user-manual_2427226026.webp)
ESO
SINFONI Pipeline User Manual
Doc:
VLT-MAN-ESO-19500-3600
Issue:
Issue 1.0
Date:
Date 2005-10-19
Page:
26 of 88
/path_raw/SCI/K/100/SINFO.2005-02-27T04:31:19.755.fits SKY_NODDING
/path_raw/SCI/K/100/SINFO.2005-02-27T04:31:46.369.fits OBJECT_NODDING
/path_raw/SCI/K/100/SINFO.2005-02-27T04:32:12.983.fits OBJECT_NODDING
/path_raw/SCI/K/100/SINFO.2005-02-27T04:32:46.418.fits SKY_NODDING
To have additional information on the instrument performance the user may want to reduce also telluric standard
star frames: those frames are caracterized by DPR.TYPE=’STD’ or DPR.TYPE=’SKY,STD’
/path_raw/STD/K/100/SINFO.2005-02-27T09:17:05.775.fits STD
/path_raw/STD/K/100/SINFO.2005-02-27T09:18:10.566.fits SKY_STD
or PSF standard star frames: those frames are caracterized by
DPR.TYPE=’PSF-CALIBRATOR’
or
DPR.TYPE=’SKY,PSF-CALIBRATOR’
/path_raw/PSF/K/100/SINFO.2004-08-23T07:35:16.597.fits PSF_CALIBRATOR
/path_raw/PSF/K/100/SINFO.2004-08-23T07:36:55.413.fits SKY_PSF_CALIBRATOR
We describe below a typical data reduction sequence using EsoRex. In this section we assume that the user
sets in the EsoRex configuration file ($HOME/.esorex/esorex.rc) the flag suppress-prefix to TRUE, so that the
pipeline product file names have standard names, with extention .fits for images and .tfits for tables. We suggest
to verify to have the flag readonly set to FALSE, if the user would like to run the same recipe several times
with EsoRex having standard names for product files. This setting allows the pipeline to overwrite previously
generated products
8
. In the following we indicate only those frames involved in the data reduction cascade,
suggesting the user to rename them according to their PRO.CATG, INS.SETUP.ID and band, and to remove the
other products after each recipe execution.
1. The user may start to generate a master dark. Raw dark frames may be put in an ASCII file, mdark_sof.
This file will look like as follows:
/path_raw/DARK/600/SINFO.2004-08-23T09:36:12.316.fits DARK
/path_raw/DARK/600/SINFO.2004-08-23T09:51:59.824.fits DARK
/path_raw/DARK/600/SINFO.2004-08-23T10:07:40.760.fits DARK
Then the user can generate the master dark with the command
esorex si_rec_mdark mdark_sof
This command will generate two files: out_bp_noise.fits (PRO.CATG=BP_MAP_HP), a hot pixel map,
and out_dark.fits (PRO.CATG=MASTER_DARK), a master dark. For convenience we indicate with
/path_cdb
the full path to a directory containing relevant data reduction products.
mv out_bp_noise.fits /path_pro/BP_MAP_HP.fits
mv out_dark.fits /path_pro/MASTER_DARK_600.fits
rm -rf out*fits *.paf *.log
8
By default installation in the EsoRex configuration file ($HOME/.esorex/esorex.rc) the flag suppress-prefix to FALSE and the flag
readonly is set to FALSE, a possible combination, in which case pipeline product filenames will have a prefix out_, an increasing a four
digit number, and extention .fits for images and .tfits for tables.