ELECTRICAL SPECIFICATION
Accuracy is expressed as:
±
(% of r no. of least significant
digits) at 23
℃±
5
℃
and <75% R.H.
Resistance (Parallel mode)
Test Frequency: 100 / 120 Hz
Accuracy
Range Maximum
Display
100 Hz/ 120Hz
Specified Note
10M
Ω
9.999M
Ω
0.6%+5
After open cal.
2000K
Ω
1999.9K
Ω
0.3%+3
After open cal.
200K
Ω
199.99K
Ω
0.3%+2 -
20K
Ω
19.999K
Ω
0.3%+2 -
2000
Ω
1999.9
Ω
0.3%+2 -
200
Ω
199.99
Ω
0.5%+3
After short cal.
20
Ω
19.999
Ω
0.6%+40
After short cal.
Test Frequency: 1K / 10K Hz
Accuracy
Range Maximum
Display
@1K Hz
@10KHz
Specified Note
10 M
Ω
9.999M
Ω
0.6%+5
2.5%+10
*N2
After open cal.
2000 K
Ω
1999.9K
Ω
0.3%+5
0.8%+10
*N2
After open cal.
200 K
Ω
199.99K
Ω
0.3%+2 0.6%+5
-
20 K
Ω
19.999K
Ω
0.3%+2 0.6%+5
-
2000
Ω
1999.9
Ω
0.3%+2 0.6%+5
-
200
Ω
199.99
Ω
0.5%+3
1.2%+25
After short cal.
20
Ω
19.999
Ω
0.6%+40
1.2%+200
After short cal.
Note:
1. DUT (Device Under Test) & test lead should be used under
properly shielding condition.
2. The accuracy is specified for parallel mode in these ranges.
11