MAW
®
User Manual Rev. 3.0
39
7.
Wipe all internal surfaces and the top of the lid with three wipes at least. Repeat until
the wipes are not discolored.
8.
Change your gloves and after a 10-minute contact time, spray sterile water and wipe
with a sterile wipe (Gamma Wipe
®
300).
9.
Inspect the device
–
if it is visually clean, consider it ready for use. If the device is
visually not clean, go to step 3 and repeat the procedure.
10.
Turn on the device (rear panel).
19 Heat optimization plates
The carry tray heat optimization plate will guarantee full contact with the dish. This gener-
ally ensures much more stable temperature conditions for cells. The plate fits in the area
under the gas hood. The plate can be taken out for cleaning and serves as a convenient way
to carry several dishes between the CO
2
incubator and the ART workstation.
Do not autoclave the carry tray. This will damage the tray, as the high tempera-
ture will make it bend out of shape.
Place the dish where it fits the pattern. There are carry trays for Nunc
®
and Falcon
®
dishes.
Use only the correct type of carry trays for your dishes.
Figure 19.1
Carry tray
20 Humidification
Use of humidification system in MAW
®
is optional and it is up to user to decide, if it is nec-
essary in each case.
21 Temperature validation
MAW, Multi-zone ART Workstation
®
is equipped with five PT-1000 Class B sensors.