21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920
E-mail: [email protected] Website:
www.escoglobal.com/ductless
USER SECTION
CHAPTER 3 – INSTALLING YOUR CABINET
3.6 OPTIONAL RETROFIT KITS
1. Cabinet is available with a transparent glass back-wall, ideal for classrooms and educational
demonstrations
2. Optional exhaust collar for exhaust ducting Ascent Opti Basic/Ascent Opti Ductless Cabinets
3. Further customisation is available upon request
3.7 INSTALLATIONS
1. Inspect your cabinet carefully. Should you find any defects, please refer to the
Freight Claim
Information on
page 6
and our
Warranty Terms and Conditions on page 7.
2. Wipe the interior and exterior of the cabinet with water or a mild household detergent.
3. (Note: This point is only applicable for SPD)Connect the cabinet to the main power supply. The fan is
automatically switched on. PLS CHECK PANEL message is displayed; please check that all LEDs
and buzzer are on for 6 seconds. Leave the blower on for 3 minutes (default warm-up time is 3
minutes; this setting can be changed though – refer to
section 4.2.1.2
).
4.
Each cabinet requires a dedicated 4A (230V) or 11.5A (115V) power socket which should not
be shared with other appliances. For some cabinets which require 2 power sockets - one for
the power outlets in the work zone (auxiliary power outlet) and the other for the cabinet blower
and electrical systems - 2 separate and dedicated power outlets should be used.