4.2.1 SETTINGS (CLOCK, TIMER, AND UNITS)
Press SET to enter the SETTINGS menu. Use the UP and DOWN button to move through the available
options. Press MENU to go back to the previous level.
4.2.1.1 SET CLOCK (TIME)
Users can set the time by increasing/decreasing the hour and minute values. The time setting is NOT
lost even when the cabinet has been switched off.
Press SET button to enter the SET TIME menu. The hour display will blink.
Press UP or DOWN arrow button to choose the correct hour value.
Press SET. The minute display will then start blinking.
Press UP or DOWN arrow button to choose the correct minute value.
Press SET button to confirm the changes.
4.2.1.2 WARM UP TIMER
Before the fan becomes fully functional, there is a time lag for the purpose of ‘warm-up’. This is to allow
some time to the sensors, fan, and the control system to stabilize. During this time contaminants in the
work zone are purged as well. The default warm-up time is 3 minutes but the users can set it to any
duration up to 15 minutes.
Press SET button to enter the WARM UP menu. The minute display will blink.
Press UP or DOWN arrow button to choose the correct minute value.
Press SET button to confirm the changes.
4.2.1.3 POST PURGE TIMER
After the user switches off the fan, there will be a period of ‘post-purge’. During this time all residual
contaminants in the work zone are purged. The default ‘post-purge’ time is 0 minutes but the users can
set it to any duration from 0 to 15 minutes. Setting it to 0 minute would effectively amount to disabling
this feature. However it is recommended that the post-purge timing be set to at least 3 minutes.
Display sequence to set time
Display sequence to set warm up time