21
6.2 IN TOP AND SIDE PANELS
Leaks can develop between the corners of two parallel top panels where they
meet two parallel side panels. Following are two methods to correct these leaks.
•
Method #1
—If the RF leak is small, the space can be filled with folded 3 oz.
copper and covered with a second strip. Remove the angle bars, place the
ends of the outer strip of copper beneath the bars, and then re bolt them
into place.
•
Method #2
—Tap along the seam with a block of wood and a hammer.
There is no need to loosen any bolts or remove any hardware for this
method. This method is also effective on 24 gauge or 26 gauge steel or
24 oz. copper enclosures.
Do not fill the space
between panels with
any material other than
what is specified here.