CAUTION
Before connecting any peripheral, verify that the total power consumption
does not exceed 350VA that is the maximum power available on insulated
sockets.
WARNING
The maximum current supplied by the
MyLab
USB ports is 500mA (for USB
2.0) and 1A (for USB 3.0). Peripherals exceeding this limit can be connected
only if powered by their external power supply.
WARNING
Epson WF-110W printer:
- the printer must be connected to the
MyLab
insulation transformer,
- do not use the printer auxiliary battery,
- the cover of the auxiliary battery shall never be removed; if the battery
cover is open, do not use the printer.