P R E P A R I N G T H E S Y S T E M
MyLab
- G E T T I N G S T A R T E D
4 - 5
GETTI
NG STARTED
Additional safety measures are:
Additional protective earth connection between the two
devices, or a safety insulation mains transformer for the
auxiliary device.
Do not connect a multiple-socket outlet or extension cord to
MyLab
.
Avoid touching the patient and the auxiliary device
simultaneously.
Additionally the IEC60601 requires control measurement of leakage currents.
The system integrator (any person connecting the medical device to other
devices) is responsible that the connections are safe.
WARNING
The system must be powered so to satisfy the electrical safety requirements,
as specified in the “Safety and Standards” manual. Esaote recommends
running a current leakage (patient and environment) test when installing in
order to check whether the applicable limits of standard EN60601-1 are not
being surpassed.
Содержание MyLab 9VET
Страница 1: ...Rev 05 February 2020 MyLab 9VET GETTING STARTED 350041300 ...
Страница 2: ......
Страница 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity ...
Страница 6: ...MyLab G E T T I N G S T A R T E D vi RED Declaration of Conformity ...
Страница 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED ...
Страница 8: ...MyLab G E T T I N G S T A R T E D viii ...
Страница 14: ...I N D E X MyLab G E T T I N G S T A R T E D xiv ...
Страница 34: ...A D D I T I O N A L I N F O R M A T I O N O N S A F E T Y MyLab G E T T I N G S T A R T E D 2 8 ...
Страница 46: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 12 ...
Страница 116: ...T E C H N I C A L S P E C I F I C A T I O N S MyLab G E T T I N G S T A R T E D 9 12 ...