Procedure
Doppler Format Optimization
1.
When required, press
TV
to activate the Tissue Velocity Doppler, when available.
2.
Adjust the Doppler gain (
Doppler
Gain key).
Rotate clockwise to increase the Doppler gain.
Rotate counterclockwise to decrease the Doppler gain.
3.
Adjust the velocity range (
VELOCITY
key).
4.
Move the zero line up or down (
BASEL
key).
5.
Adjust the frequency (
FREQ
key).
6.
If necessary, adjust the Doppler steer (
D-STEER
or
STEER
key) first
activating the 2D real time (
PLEX
key).
7.
If necessary, press the
ANGLE
key to align the angle vector with the
flow direction.
8.
If necessary, change the size of the sample volume (
SV SIZE
key).
9.
If necessary, press the
REVERSE
key to reverse the flow direction.
10.
Adjust the 2D real time display (
B-FORMAT
key).
11.
If necessary, press
B-REF
to view the Doppler trace at full screen.
12.
Rotate the
POWER
key to change the transmitted power, using the
minimum power compatible with a diagnostic level of the images.
Performing an Exam
Содержание mylab 25 gold
Страница 2: ...QG_ML25 30GOLD RevC 3 19 03 2009 10 10 Pagina 2 Colori compositi C M Y CM MY CY CMY K ...
Страница 16: ...QG_ML25 30GOLD RevC 3 19 03 2009 10 10 Pagina 16 Colori compositi C M Y CM MY CY CMY K Your Notes ...
Страница 38: ...QG_ML25 30GOLD RevC 3 19 03 2009 10 10 Pagina 38 Colori compositi C M Y CM MY CY CMY K Your Notes ...
Страница 50: ...QG_ML25 30GOLD RevC 3 19 03 2009 10 10 Pagina 50 Colori compositi C M Y CM MY CY CMY K Your Notes ...
Страница 52: ...QG_ML25 30GOLD RevC 3 19 03 2009 10 10 Pagina 52 Colori compositi C M Y CM MY CY CMY K Your Notes ...
Страница 58: ...QG_ML25 30GOLD RevC 3 19 03 2009 10 10 Pagina 58 Colori compositi C M Y CM MY CY CMY K Your Notes ...