Your application was analyzed and validated by the erlab® laboratory as part of the erlab
Safety Program (E.S.P.). During this validation, the erlab® laboratory used the information
provided to determine the chemical that must be the subject of the molecular filter satu-
ration test. You have access to this information at any time on your Valipass® certificate.
MANUALLY DETECTING FILTER SATURATION
Verify the chemical for which you will
carry out the fiter saturation test with
a reactive tube
To carry out a manual filter saturation test, you need the following equipment:
- a pump (a simple plastic pump A is provided, For more accurate result use a
professional pump type B from Draeger, Gastec, Kitagawa, RAE, etc…)
- a flexible silicone tube (provided)
- the chemical to be tested (not provided)
- an appropriate color coded tube for the chemical (Dreager, Gastec, Kitagawa, RAE, etc…)
Turn on the Captair® Flex™ fume hood
Turn on the ventlation
A
B