Event 7:
Pre-warning alarm maintenance
Event conditions:
If the Timer™’s pre-warning alarm is activated, this alarm is triggered when the counter’s
value is less than the pre-warning value.
FILTER TEST
H / MN
Display screen
Alarm type
Solution :
Press the navigation and confirmation buttons simultaneously and release them to deacti-
vate the alarm.
Consider checking whether or not the filter is saturated.
Refer to the maintenance section of this manual.
Event 8:
Maintenance
Event conditions:
The value of the Timer™ is equal to zero.
FILTER
TEST
Display screen
Alarm type
Solution :
Press the navigation and confirmation buttons simultaneously and release them to deacti-
vate the alarm.
Consider checking whether or not the filter is saturated.
Refer to the maintenance section of this manual.
Yellow indicator light on, blinking
Intermittent audible alarm
Red indicator light on, constant
Continuous audible alarm