User Manual
TM
11
Event 2:
Fan out of order
Event conditions:
The rotation speed of the fan is less than 700 RPM.
FAN
OUT OF ORDER
Display screen
Alarm type
Solution :
Contact your maintenance service as quickly as possible.
Solution :
Press the navigation and confirmation buttons simultaneously and release them to deacti-
vate the alarm.
The red indicator light will stay on.
The concentration at the exhaust of a filter nearing its saturation point can build up very
quickly. Consider replacing the filter.
Event 3:
Solvent detected in the sampling chamber
Event conditions:
The detection limit of the Molecode™ S unit has been reached.
FILTER
NEAR END OF LIFE
Display screen
Alarm type
Red indicator light on, constant
Continuous audible alarm
Yellow indicator light on, constant
Intermittent audible alarm